ODiSI 6000 Series Optical Distributed Sensor Interrogator for High-Definition Fiber Optic Strain and Temperature Sensing
Multichannel Rayleigh-based HD-FOS interrogator measuring distributed strain and temperature with gage pitch down to 0.65 mm on up to 8 channels.
- Number of channels
- 1, 2, 4 or 8 channels
- Maximum sensor length per channel, standard mode
- 20 m
- Maximum sensor length per channel, extended range
- 100 m
- Standoff cable length
- 10, 50, 100, 150 or 200 m
Overview
The ODiSI 6000 Series from Luna is a family of multichannel high-definition fiber optic sensing (HD-FOS) interrogators. The system works with sensors made from unaltered optical fiber and measures Rayleigh backscatter to acquire distributed strain and temperature data, with a gage pitch as fine as 0.65 mm. This produces thousands of measurement points per meter of sensing fiber and allows strain contours or continuous temperature profiles to be mapped in detail.
The series is offered in two model lines. The ODiSI 610x models (6101/6102/6104/6108) support gage pitches of 0.65 mm, 1.3 mm, 2.6 mm and 5.2 mm, while the ODiSI 600x models (6001/6002/6004/6008) use a 5.2 mm gage pitch. Each is available with 1, 2, 4 or 8 channels. With Standard Remote Modules each channel supports an HD sensor up to 20 m or a continuous fiber grating (CFG) sensor up to 15 m; optional Extended Range Remote Modules extend this to 100 m and enable HD-SC strain-compensated temperature sensors.
Sensors
HD strain sensors use polyimide-coated fiber, are calibrated with NIST-traceable fixtures and are supplied with scaling coefficients and a reference key for plug-and-play identification. CFG sensors, with fiber Bragg gratings inscribed continuously along the fiber, provide increased tolerance to vibration and improved dynamic performance. The sensors are passive and dielectric and can be bonded to curved surfaces or embedded within materials.
Software and integration
The system is a complete data acquisition solution: the ODiSI application software configures sensors, measurements and triggering, and logs, replays and exports data. Data can be streamed in real time over Ethernet in a JSON-based format, and ODiSI Remote software for Windows and Linux provides remote control and viewing. Integration features include IEEE 1588 PTP time synchronization, a TTL sync output and start/stop triggering. Optional 3DViz software maps real-time or saved data onto 3D CAD models or 2D images.
The ODiSI 6000 Series has been discontinued by the manufacturer; the replacement product is the ODiSI 7100 Series.
Models in this series
| Model | Description | |
|---|---|---|
| ODiSI 6001 |
Luna ODiSI 6001 Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
ODiSI 600x-class distributed fiber optic strain and temperature interrogator from the ODiSI 6000 Series, using a standard 5.2 mm gage pitch. |
Details → |
| ODiSI 6004 |
Luna ODiSI 6004 4-Channel Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
Four-channel ODiSI 600x distributed sensor interrogator that measures strain and temperature along optical fiber with a 5.2 mm gage pitch. |
Details → |
| ODiSI 6101 |
Luna ODiSI 6101 Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
ODiSI 610x-family HD-FOS interrogator measuring distributed strain and temperature over Rayleigh-backscatter fiber with gage pitch down to 0.65 mm. |
Details → |
| ODiSI 6002 |
ODiSI 6000 Series ODiSI 6002 Two-Channel Optical Distributed Sensor Interrogator
Two-channel Rayleigh backscatter interrogator for distributed strain and temperature sensing with a 5.2 mm gage pitch. |
Details → |
| ODiSI 6008 |
ODiSI 6008 8-Channel Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
Eight-channel ODiSI 600x distributed sensing instrument that measures strain and temperature along optical fiber at a 5.2 mm gage pitch. |
Details → |
| ODiSI 6102 |
ODiSI 6102 2-Channel Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
Two-channel ODiSI 610x high-definition fiber optic sensing interrogator for distributed strain and temperature with gage pitch down to 0.65 mm. |
Details → |
| ODiSI 6104 |
ODiSI 6104 4-Channel Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
Four-channel ODiSI 6000 Series interrogator for high-definition distributed strain and temperature sensing with gage pitch down to 0.65 mm. |
Details → |
| ODiSI 6108 |
ODiSI 6108 8-Channel Optical Distributed Sensor Interrogator
Eight-channel ODiSI 6000 Series HD-FOS interrogator for distributed strain and temperature sensing with gage pitch down to 0.65 mm. |
Details → |
Key Features
- High-definition fiber optic sensing (HD-FOS) with sub-millimeter gage pitch
- Multichannel measurements with up to 8 channels per system
- Up to 100 m of high-definition sensing fiber per channel with Extended Range Remote Modules
- Complete system measures, acquires, logs and streams real-time data
- Flexible, lightweight and easy-to-install fiber optic sensors
- Passive, corrosion resistant, dielectric sensors for bends, corners and embedded installations
- Long sensor life with no drift or recalibration, cycle counts >10^7
- Continuous optical alignment and calibration with every scan
- Strain sensors calibrated with NIST-traceable fixtures
- Optional strain-compensated temperature mode
- Compatible with continuous fiber grating (CFG) sensors for improved dynamic performance
- IEEE 1588 PTP network time synchronization
- JSON-based digital streaming data format
- Sync output TTL pulse and start/stop triggering
- Automatic plug-and-play sensor identification
- Real-time streaming of data over Ethernet
- ODiSI Remote software for Windows and Linux
Specifications
| Number of channels | 1, 2, 4 or 8 channels |
|---|---|
| Maximum sensor length per channel, standard mode | 20 m |
| Maximum sensor length per channel, extended range |
100 m
Sensor lengths longer than 20 m, as well as HD-SC strain-compensated sensors, require optional Extended Range Remote Modules |
| Standoff cable length |
10, 50, 100, 150 or 200 m
Standard lengths 10, 50 and 100 m; two cables can be combined to form 150 and 200 m |
| Maximum patch cord length |
Up to 19 m (depending on sensor length)
Patch cord plus active sensor fiber must not exceed 20 m (Standard) or 100 m (Extended Range) per channel |
| Data interface | Ethernet |
|
Gage pitch (HD sensors)
ODiSI 610x |
0.65 mm, 1.3 mm, 2.6 mm, 5.2 mm
Gage pitch is the distance between centerpoints of consecutive gages |
|---|---|
|
Gage pitch (HD sensors)
ODiSI 600x |
5.2 mm |
|
Gage pitch (CFG sensors)
ODiSI 610x, ODiSI 600x |
6 mm
CFG mode requires Luna CFG sensors |
| Gages per meter of sensor | 1,538 (0.65 mm); 768 (1.3 mm); 384 (2.6 mm); 192 (5.2 mm); 166 (6 mm CFG) |
|
Max measurement rate
ODiSI 610x |
250 Hz |
|
Max measurement rate
ODiSI 600x |
30 Hz |
|
0.65 mm gage pitch — 2.5 m / 5 m / 10 m / 20 m mode
ODiSI 610x |
62.5 Hz / 40 Hz / 25 Hz / 12.5 Hz
Rates are aggregate; divide by number of active channels for per-channel rate. Not available in 50 m and 100 m modes |
|---|---|
|
1.3 mm gage pitch — 2.5 m / 5 m / 10 m / 20 m / 50 m mode
ODiSI 610x |
125 Hz / 80 Hz / 50 Hz / 25 Hz / 10 Hz
Rates are aggregate; divide by number of active channels. Not available in 100 m mode |
|
2.6 mm gage pitch — 2.5 m / 5 m / 10 m / 20 m / 50 m / 100 m mode
ODiSI 610x |
250 Hz / 160 Hz / 100 Hz / 50 Hz / 20 Hz / 10 Hz
Rates are aggregate; divide by number of active channels |
|
5.2 mm gage pitch — 2.5 m / 5 m / 10 m / 20 m / 50 m / 100 m mode
ODiSI 610x |
250 Hz / 160 Hz / 100 Hz / 50 Hz / 20 Hz / 10 Hz
Rates are aggregate; divide by number of active channels |
|
5.2 mm gage pitch — 2.5 m / 5 m / 10 m / 20 m / 50 m / 100 m mode
ODiSI 600x |
30 Hz / 20 Hz / 12.5 Hz / 6.3 Hz / 2.5 Hz / 1 Hz
Rates are aggregate; divide by number of active channels |
|
6 mm CFG gage pitch — 2.5 m / 5 m / 10 m / 20 m mode
ODiSI 610x, ODiSI 600x |
62.5 Hz / 40 Hz / 25 Hz / 12.5 Hz
Rates are aggregate; divide by number of active channels. Not available in 50 m and 100 m modes |
| Strain measurement range | ±15,000 µϵ |
|---|---|
| Resolution | 0.1 µϵ |
| Instrument accuracy | ±1 µϵ |
|
System (instrument and sensor) accuracy, 0.65 mm gage pitch
ODiSI 610x |
±25 µϵ
Compared to NIST-traceable extensometer measurements; average of 150 measurements at each of seven strain increments |
| System (instrument and sensor) accuracy, 1.3 / 2.6 / 5.2 / 6 mm (CFG) gage pitch |
±30 µϵ
CFG accuracy specifications are preliminary |
| Measurement uncertainty, 0.65 / 1.3 / 2.6 / 5.2 / 6 mm (CFG) gage pitch |
±5 µϵ / ±4 µϵ / ±2 µϵ / ±2 µϵ / ±5 µϵ
Twice the standard deviation of 1000 measurements, at zero and across full strain range; includes interrogator and Luna sensors |
| Dynamic loading rate, 0.65 / 1.3 / 2.6 / 5.2 / 6 mm (CFG) gage pitch | 1 Hz / 2.5 Hz / 5 Hz / 5 Hz / tbd |
| Temperature measurement range |
-40 to 200 °C (with standard Luna sensors)
With standard temperature sensors (HD6T and HD6SCT); wider range available with other sensor fiber |
|---|---|
| Temperature measurement resolution | 0.1 °C |
| Temperature measurement uncertainty, 0.65 / 1.3 / 2.6 / 5.2 / 6 mm (CFG) gage pitch |
±2.2 °C / ±1.6 °C / ±0.6 °C / ±0.6 °C / ±1.0 °C
Twice the standard deviation of 1000 measurements |
| Gage pitch |
1.04 cm
Requires Luna HD-SC temperature sensors and Extended Range Remote Modules |
|---|---|
| Measurement rate (all lengths) | 2.5 Hz |
| Resolution | 0.1 °C |
| Measurement uncertainty | ±0.9 °C |
| Strain compensation |
Over an applied strain range of 0 to 1800 µϵ, measurement accuracy is ± 2.8 °C
RMS error with HD-SC sensor under axial strain; includes effects of the interrogator |
| Sensor lengths | 1 to 5 m |
| Class 1 laser | <10 mW |
|---|---|
| Operating temperature range — mainframe and controller | 5 to 40 °C |
| Operating temperature range — standoff cable and remote module | 5 to 60 °C |
| Storage temperature range | 0 to 40 °C |
| Operating and storage relative humidity |
10 to 90 % RH
Non-condensing |
| Operating altitude | -15 to 3,000 m |
| Dimensions — mainframe (W x D x H) | 34 x 35 x 11 cm |
| Dimensions — laptop controller (W x D x H) | 36 x 24 x 1.7 cm |
| Weight (mainframe and controller) | 17 lb (7.8 kg) |
| Power (mainframe and controller) | 160 W |
Applications
- High fidelity model verification
- FE model validation with 2D and 3D strain fields
- Temperature mapping for batteries
- Bonding and welding quality improvement
- Multi-material joining evaluation
- Embedded sensors for smart parts
- Strain characterization of new materials and complex structures
- In-situ process temperature profiling
- Crack and defect detection in critical structures
- Composite materials testing
Options & Accessories
| Option | Description |
|---|---|
| OP06150 | Extended Range Remote Module — Remote module supporting extended sensor lengths of 50 m and 100 m, in addition to standard sensor lengths and strain-compensated temperature (HD-SC) sensors. |
| OPT06100-R | Rack Mount Controller Option — Rack mount (1U) instrument controller (substituted for laptop). |
| OPT06020 | USB-3106 Analog Output — 16-channel analog output module for ODiSI (voltage or current outputs). |
| OPT06125 | 3Dviz - 3D/2D Visualization Software — Visualization software for mapping real-time or saved ODiSI data onto 3D models and 2D images. Compatible with Windows OS. |
| OPT06123 | 3Dviz - 2D Visualization Software — Visualization software for mapping real-time or saved ODiSI data onto 2D images. Compatible with Windows OS. |
| HD6SXXLC300P | HD strain sensor — HD strain sensor, LC2+ connector, polyimide coating, 300 °C, length = XX m (1 to 100 m) |
| HD6TXXLC220P | HD temperature sensor — HD temperature sensors with PTFE sleeving, length = XX m (1 to 5 m) |
| HD6SCTXXLC220P | HD strain-compensated temperature sensor — HD strain-compensated temperature sensors with PTFE sleeving, 200 °C, length = XX m (1 to 5 m) |
| HDCFGXXXLCP | HD continuous fiber grating (CFG) sensor — HD continuous fiber grating (CFG) sensors, length = XXX m (1 to 15 m) |
| FOSAPPKIT | Fiber Optic Sensor Application Kit — Materials (epoxy, tools, etc.) required to install 10 m of sensing fiber |