Luna ODiSI 6101 Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
ODiSI 610x-family HD-FOS interrogator measuring distributed strain and temperature over Rayleigh-backscatter fiber with gage pitch down to 0.65 mm.
- Number of channels
- 1, 2, 4 or 8 channels
- Maximum sensor length per channel — standard mode
- 20 m
- Maximum sensor length per channel — extended range
- 100 m
- Standoff cable length
- 10, 50, 100, 150 or 200 m
Overview
The ODiSI 6101 belongs to the ODiSI 610x models of the Luna ODiSI 6000 Series of optical distributed sensor interrogators. Luna lists this product as discontinued and names the ODiSI 7100 Series as its replacement.
The ODiSI 6000 Series instruments are high-definition fiber optic sensing (HD-FOS) interrogators. They work with sensors made from unaltered optical fiber and measure Rayleigh backscatter. This lets them record strain and temperature at thousands of points per meter of sensing fiber, mapping the strain contour of a structure under test or the continuous temperature profile of a process.
Measurement modes
The ODiSI 610x models can be set to measure strain or temperature with a gage pitch of 0.65 mm, 1.3 mm, 2.6 mm or 5.2 mm. They also work with continuous fiber grating (CFG) sensors, which tolerate vibration better and give improved dynamic performance. With the optional Extended Range Remote Modules, a channel supports sensors up to 100 m long and a strain-compensated temperature mode for surface-mounted and embedded HD-SC sensors.
Software and integration
The system comes with an instrument controller running the ODiSI application software. The software identifies sensors automatically and handles configuration, triggering, data logging, replay and export. Data can be streamed in real time over Ethernet in a JSON-based format. The system supports IEEE 1588 PTP time synchronization, a TTL sync output and start/stop triggering. ODiSI Remote software for Windows and Linux provides remote control and viewing, and the optional 3DViz software maps data onto 3D CAD models or 2D images.
Key Features
- High-definition fiber optic sensing (HD-FOS) with sub-millimeter sampling resolution
- Gage pitch of 0.65 mm, 1.3 mm, 2.6 mm or 5.2 mm on ODiSI 610x models
- Maximum measurement rate of 250 Hz on ODiSI 610x models
- Up to 100 m of high-definition sensing per channel with Extended Range Remote Module
- Complete system measures, acquires, logs and communicates real-time data
- Compatible with HD strain, HD temperature, HD-SC strain-compensated temperature and CFG sensors
- Passive, corrosion resistant, dielectric sensors for bends, corners and embedded locations
- Long sensor life with no drift or recalibration, cycle counts >10^7
- Continuous optical alignment and calibration with every scan
- Strain sensors calibrated with NIST-traceable fixtures
- IEEE 1588 PTP network time synchronization
- JSON-based digital streaming data format
- Sync output TTL pulse
- Start and stop triggering functions
- Real-time remote access via ODiSI Remote software for Windows and Linux
Specifications
| Number of channels |
1, 2, 4 or 8 channels
Ordering designation ODiSI 6101/2/4/8
|
|---|---|
| Maximum sensor length per channel — standard mode | 20 m |
| Maximum sensor length per channel — extended range |
100 m
Sensor lengths longer than 20 m, as well as HD-SC strain-compensated sensors, require optional Extended Range Remote Modules
|
| Standoff cable length |
10, 50, 100, 150 or 200 m
Standard lengths of 10, 50 and 100 m; two cables can be combined to form standoff lengths of 150 and 200 m
|
| Maximum patch cord length |
Up to 19 m (depending on sensor length)
Total patch cord plus active sensor fiber must not exceed 20 m (Standard) or 100 m (Extended Range) per channel
|
| Data interface | Ethernet |
|
Gage pitch (HD Rayleigh sensors)
ODiSI 610x
|
0.65 mm, 1.3 mm, 2.6 mm, 5.2 mm
Gage pitch is the distance between centerpoints of consecutive gages
|
|---|---|
|
Gages per meter of sensor (HD)
ODiSI 610x
|
1,538 (0.65 mm), 768 (1.3 mm), 384 (2.6 mm), 192 (5.2 mm) |
| Gage pitch (CFG sensors) |
6 mm
CFG mode requires Luna CFG sensors; 166 gages per meter
|
|
Max measurement rate
ODiSI 610x
|
250 Hz |
|
Measurement rate — 2.5 m mode
ODiSI 610x
|
62.5 Hz (0.65 mm), 125 Hz (1.3 mm), 250 Hz (2.6 mm), 250 Hz (5.2 mm); CFG: 62.5 Hz
Rates are aggregate; divide by number of active channels to determine the per-channel rate
|
|
Measurement rate — 5 m mode
ODiSI 610x
|
40 Hz (0.65 mm), 80 Hz (1.3 mm), 160 Hz (2.6 mm), 160 Hz (5.2 mm); CFG: 40 Hz
Aggregate rates
|
|
Measurement rate — 10 m mode
ODiSI 610x
|
25 Hz (0.65 mm), 50 Hz (1.3 mm), 100 Hz (2.6 mm), 100 Hz (5.2 mm); CFG: 25 Hz
Aggregate rates
|
|
Measurement rate — 20 m mode
ODiSI 610x
|
12.5 Hz (0.65 mm), 25 Hz (1.3 mm), 50 Hz (2.6 mm), 50 Hz (5.2 mm); CFG: 12.5 Hz
Aggregate rates
|
|
Measurement rate — 50 m mode
ODiSI 610x
|
10 Hz (1.3 mm), 20 Hz (2.6 mm), 20 Hz (5.2 mm)
Not available at 0.65 mm or with CFG sensors; aggregate rates
|
|
Measurement rate — 100 m mode
ODiSI 610x
|
10 Hz (2.6 mm), 10 Hz (5.2 mm)
Not available at 0.65 mm, 1.3 mm or with CFG sensors; aggregate rates
|
| Strain measurement range | ±15,000 µϵ |
|---|---|
| Resolution | 0.1 µϵ |
| Instrument accuracy | ±1 µϵ |
|
System (instrument and sensor) accuracy
ODiSI 610x
|
±25 µϵ (0.65 mm), ±30 µϵ (1.3 mm), ±30 µϵ (2.6 mm), ±30 µϵ (5.2 mm); CFG: ±30 µϵ
Compared to NIST-traceable extensometer measurements; CFG specifications are preliminary
|
|
Measurement uncertainty
ODiSI 610x
|
±5 µϵ (0.65 mm), ±4 µϵ (1.3 mm), ±2 µϵ (2.6 mm), ±2 µϵ (5.2 mm); CFG: ±5 µϵ
At zero and across the full strain range; twice the standard deviation of 1000 measurements
|
|
Dynamic loading rate
ODiSI 610x
|
1 Hz (0.65 mm), 2.5 Hz (1.3 mm), 5 Hz (2.6 mm), 5 Hz (5.2 mm); CFG: tbd |
| Temperature measurement range |
-40 to 200 °C (with standard Luna sensors)
With standard temperature sensors (HD6T and HD6SCT); wider range available with other sensor fiber
|
|---|---|
| Temperature measurement resolution | 0.1 °C |
|
Temperature measurement uncertainty
ODiSI 610x
|
±2.2 °C (0.65 mm), ±1.6 °C (1.3 mm), ±0.6 °C (2.6 mm), ±0.6 °C (5.2 mm); CFG: ±1.0 °C |
| Gage pitch |
1.04 cm
Requires Luna HD-SC temperature sensors and Extended Range Remote Modules
|
|---|---|
| Measurement rate (all lengths) | 2.5 Hz |
| Resolution | 0.1 °C |
| Measurement uncertainty | ±0.9 °C |
| Strain compensation |
Over an applied strain range of 0 to 1800 µϵ, measurement accuracy is ± 2.8 °C
RMS error with the HD-SC sensor under axial strain; includes effects of the ODiSI interrogator
|
| Class 1 laser | <10 mW |
|---|---|
| Operating temperature range — mainframe and controller | 5 to 40 °C |
| Operating temperature range — standoff cable and remote module | 5 to 60 °C |
| Storage temperature range | 0 to 40 °C |
| Operating and storage relative humidity |
10 to 90 % RH
Non-condensing
|
| Operating altitude | -15 to 3,000 m |
| Dimensions — mainframe (W x D x H) | 34 x 35 x 11 cm |
| Dimensions — laptop controller (W x D x H) | 36 x 24 x 1.7 cm |
| Weight (mainframe and controller) | 17 lb (7.8 kg) |
| Power (mainframe and controller) | 160 W |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- High fidelity model verification
- Temperature mapping for batteries
- Improving bonding and welding quality
- Embedded sensors for smart parts
- Strain characterization of new materials and complex structures
- In-situ temperature profiling of critical processes
- Two- and three-dimensional strain field measurement for FE model validation
- Evaluation of multi-material joining
- Detection of small cracks and defects in critical structures
- Strain measurement on and in composite materials
Included
- ODiSI mainframe
- Instrument controller laptop
- ODiSI application software
- One standoff cable for each channel
- One standard length remote module for each channel
Options & Accessories
| Option | Description |
|---|---|
| OP06150 | Extended Range Remote Module — Remote module supporting extended sensor lengths of 50 m and 100 m, in addition to standard sensor lengths and strain-compensated temperature (HD-SC) sensors. |
| OPT06100-R | Rack Mount Controller Option — Rack mount (1U) instrument controller (substituted for laptop). |
| OPT06020 | USB-3106 Analog Output — 16-channel analog output module for ODiSI (voltage or current outputs). |
| OPT06125 | 3Dviz - 3D/2D Visualization Software — Visualization software for mapping real-time or saved ODiSI data onto 3D models and 2D images. Compatible with Windows OS. |
| OPT06123 | 3Dviz - 2D Visualization Software — Visualization software for mapping real-time or saved ODiSI data onto 2D images. Compatible with Windows OS. |
| HD6SXXLC300P | HD strain sensor — HD strain sensor, LC2+ connector, polyimide coating, 300 °C, length = XX m (1 to 100 m) |
| HD6TXXLC220P | HD temperature sensor — HD temperature sensors with PTFE sleeving, length = XX m (1 to 5 m) |
| HD6SCTXXLC220P | HD strain-compensated temperature sensor — HD strain-compensated temperature sensors with PTFE sleeving, 200 °C, length = XX m (1 to 5 m) |
| HDCFGXXXLCP | HD continuous fiber grating (CFG) sensor — HD continuous fiber grating (CFG) sensors, length = XXX m (1 to 15 m) |
| FOSAPPKIT | Fiber Optic Sensor Application Kit — Materials (epoxy, tools, etc.) required to install 10 m of sensing fiber |