CX300 ComXpert Communications Service Monitor
Portable communications service monitor for analog and digital radio testing from 100 kHz to 6 GHz, with optional Auto-Test alignment.
- Display size
- 10 in (25.4 cm)
Overview
The VIAVI CX300 ComXpert is a communications service monitor for radio technicians. It comes in a small, ruggedized, field-portable form factor and supports both analog and digital radio testing.
The CX300 has a large color touchscreen. Its user interface uses multiple tabs, much like an internet browser, so the technician can switch between test setups without reloading settings. Color-coded meters look like swept analog meters and also show a digital readout. Individual and overall pass/fail flags support go/no-go testing.
The VIAVI Auto-Test applications are available as an option. They run automated test and alignment routines that follow each radio OEM's maintenance specifications.
Data management and remote control
Test results from manual and automated tests can be saved and managed with VIAVI StrataSync, a cloud-based system for asset configuration and test data management. The CX300's wireless connectivity lets it upload results to the cloud and receive software updates. The free VIAVI Smart Access Anywhere app, available for Windows, Android and iOS, provides full remote control of the test set.
Key Features
- Frequency range from 100 kHz to 6 GHz
- Multiple tab user interface, similar to an internet browser
- Intelligent color-coded meters for at-a-glance test verification
- VIAVI Auto-Test applications for all major LMR and PMR radios
- VIAVI StrataSync integration offers asset, configuration, and test data management
- Remotely control the test set using Windows, Android, and iOS with Smart Access Anywhere
- 10 in color touchscreen display
- Digital modes: P25, P25 Phase 2, DMR, NXDN, TETRA
- Built-in spectrum analyzer, oscilloscope, audio analyzer and vector network analyzer
Specifications
| Display size | 10 in (25.4 cm) |
|---|
| Accuracy |
0.02 ppm
0°C to 50°C
|
|---|---|
| Aging | ±0.1 ppm/year |
| Warm-up time | 3 minutes: within ±0.01 ppm |
| Accuracy with GPS | ±25 ppb (GPS Lock); ±50 ppb (Hold over 72 hours) |
| External reference | 10 MHz |
| Frequency range |
100 kHz to 3GHz
Standard
|
|---|---|
|
Frequency range
CX300-F6GHz
|
3 GHz to 6 GHz |
| Frequency resolution | 1 Hz |
| Output level, RF Duplex Port | -140 dBm to -30 dBm (10 MHz to 1 GHz); -37 dBm for AM and Complex modulation |
| Output level, RF Output Port | -130 dBm to +17 dBm (10 MHz to 1 GHz); +10 dBm for AM and Complex modulation |
| Output level resolution | 0.1 dB |
| Output level accuracy | ±1.0 dB (output level >-120 dBm, 1 MHz to 6 GHz); ±2.0 dB (output level >-130 dBm, 1 MHz to 6 GHz); ±1.0 dB typical |
| Bandwidth | 100 MHz |
| VSWR, RF Duplex Port | <1.1 (1 MHz to 1 GHz); <1.2 (1 GHz to 6 GHz) |
| VSWR, RF Output Port | <1.4 (1 MHz to 1 GHz); <1.5 (1 GHz to 6 GHz) |
| Phase noise |
-112 dBc/Hz at 10 kHz offset at 500 MHz; -110 dBc/Hz at 10 kHz offset at 1000 MHz
Frequency ≥1 MHz and Level ≤+10 dBm
|
| Harmonics |
-35 dBc
Frequency ≥1 MHz and Level ≤+10 dBm
|
| Non-harmonics |
-45 dBc
Frequency ≥1 MHz and Level ≤+10 dBm
|
| Residual AM | <0.1% RMS |
| Residual FM | <3 Hz RMS 300 Hz to 3 kHz |
| Modes | AM, FM, PM, SSB |
|---|---|
| Modulation frequency range | 20 Hz to 20 kHz |
| Distortion | <1% THD |
| AM range |
0% to 100%
Resolution 0.1%; accuracy (internal source) ≤±5% of settings
|
| FM range |
0 Hz to 100 kHz
Resolution 1 Hz; accuracy (internal source) ≤±2.5% of setting with frequency response of ±0.5 dB 20 Hz to 10 kHz
|
| PM range |
0 rad to 6.3 rad
Resolution 0.1 rad; accuracy <±2.5% of setting with frequency response of ±0.5 dB 20 Hz to 10 kHz
|
| SSB carrier suppression | >70 dB |
| SSB sideband suppression | >60 dB |
| Number of internal modulation sources | 3 |
| Source waveforms | Sine, Square, Triangle, Ramp, DTMF, DCS, CTCSS, Tone Remote, Tone Sequential, Two-Tone Sequential |
| CTCSS tone | Tone 1 (67) to Tone 50 (254.1) Hz |
| Frequency range |
9 kHz to 3 GHz
Standard
|
|---|---|
|
Frequency range
CX300-F6GHz
|
3 GHz to 6 GHz |
| Maximum input level, RF Input Port | +27 dBm (500 mW) max preamp and frequency ≥1 MHz; +13 dBm (20 mW) max preamp on or frequency <1 MHz |
| Maximum input level, RF Duplex Port |
+47 dBm (50 Watts) continuous, +<35°C; +51 dBm (125 Watts) Cyclical (Max “ON” of 30 sec and Min “OFF” for 90 sec) for power levels >50 Watts
Shutdown: alarm sounds (no auto shutdown)
|
| VSWR, RF Duplex Port | ≤1.2 (100 kHz to 1 GHz) |
| VSWR, RF Input Port |
≤1.6 (100 kHz to 1 GHz)
with 10 dB input attenuation
|
| Spurious response, input related | ≤-65 dBc typical |
| Spurious response, non-input related | ≤-95 dBm typical |
| Phase noise | -112 dBc/Hz at 10 kHz offset at 500 MHz; -110 dBc/Hz at 10 kHz offset at 1000 MHz |
| Dynamic range | 2/3 * (TOI-DANL) = 109 dB |
| TOI | +20 dBm (0 atten), >+1 dBm (preamp), 1 MHz to 1 GHz |
| DANL | 900 MHz: <-146 dBm (0 dB attenuation), -162 dBm (preamp); 1000 MHz: <-142 dBm (0 dB attenuation), <-160 dBm (preamp) |
| Analog sensitivity |
10 dB SINAD, <-105 dBm with preamp
300 Hz to 3 kHz audio filter, 2.5 kHz FM deviation, 12.5 kHz IF BW
|
| Bandwidth | 100 MHz (wideband VSA), 8 MHz (narrowband) |
| RF bandpass filters (IF filters) | 250 Hz, 3 kHz, 5 kHz, 6.25 kHz, 8.33 kHz, 10 kHz, 12.5 kHz, 25 kHz, 30 kHz, 50 kHz, 100 kHz, 230 kHz, 300 kHz, 1 MHz, 3 MHz, 5 MHz, 10 MHz, 20 MHz |
| Frequency range |
100 kHz to 3 GHz
Standard; 3 GHz to 6 GHz with CX300-F6GHz
|
|---|---|
| Measurement modes | RMS, average RMS, minimum, maximum |
| Level, RF Duplex Port | -20 dBm to +51 dBm |
| Level, RF Input Port | -60 dBm to +10 dBm |
| Accuracy, RF Duplex Port |
±0.4 dB (1 MHz to 1 GHz); ±0.6 dB (1 GHz to 6 GHz)
Accuracy after normalizing at the measurement frequency
|
| Accuracy, RF Input Port |
±0.8 dB (1 MHz to 1 GHz), ±0.9 dB (1 GHz to 6 GHz)
Accuracy after normalizing at the measurement frequency
|
| Frequency range |
100 kHz to 3 GHz
Standard; 3 GHz to 6 GHz with CX300-F6GHz
|
|---|---|
| Resolution | 1 Hz |
| Input level range, RF Duplex Port | -20 dBm to 51 dBm |
| Input level range, RF Input Port | -60 dBm to +17 dBm (-80 dBm to -20 dBm w/pre-amp) |
| FM modes | RMS, +PK, -PK, ±PK/2 |
|---|---|
| FM measurement range | 0 Hz to 75 kHz |
| FM accuracy | ±1.0% for rate ≥1.5 kHz and ≤3 kHz; ±2.0% otherwise |
| Residual FM | ≤3 Hz (300 Hz to 3 kHz) and frequency <1 GHz |
| AM measurement range | 0% to 100% |
| AF frequency range | 10 Hz to 20 kHz |
| PM range |
0 radians to 6.3 radians
Resolution 0.01 rad for ≤5 rad, 0.1 rad for >5 rad
|
| SSB modes | SSB-USB, SSB-LSB |
| SINAD meter range |
0 dB to 63 dB
50 Hz to 10 kHz; accuracy <±1 dB
|
|---|---|
| Distortion meter range |
0% to 100%
50 Hz to 10 kHz
|
| AF tones analyzer modes | DTMF, DCS, CTCSS, Two-Tone, Tone Sequential, Tone Remote |
| Audio level meter input impedance | 100 K Ω, 600 Ω, 300 Ω |
| Audio analyzer frequency range | DC to 100 kHz |
| Audio filters, lowpass | 300 Hz, 3 kHz, 3.4 kHz, 5 kHz, 15 kHz, 20 kHz, 40 kHz, TIA 3 kHz, TIA 15 kHz |
| Audio filters, highpass | 50 Hz, 300 Hz, TIA 50 Hz, TIA 300 Hz |
| Audio filters, other | C-MSG, CCITT |
| Span | 2 kHz to 8 MHz |
|---|---|
| IF bandwidth | 10 MHz |
| RBW | 1 Hz to 50 kHz |
| FFT windows | Flat top, rectangular, Hamming, Hanning, Blackman-Harris |
| Frequency range |
9 kHz to 3 GHz
Standard
|
|---|---|
|
Frequency range
CX300-F6GHz
|
3 GHz to 6 GHz |
| RBW range | 25 Hz to 6 MHz |
| VBW range | 5 Hz to 6 MHz |
| Sweep time range | 0.4 ms to 1000 s |
| Spurious free dynamic range | ≥80 dB |
| DANL | <-142 dBm (0 atten), <-162 dBm (preamp) |
| Zero span sweep time range | 24 µs to 200 s |
| Tracking generator output ports | RF Output Port, RF Duplex Port |
| I/Q recorder sample length | 4 Mega Samples |
| Output impedance | <4 Ω |
|---|---|
| Max output current | 100 mA |
| Frequency range |
0 Hz to 100 kHz
Resolution 0.1 Hz
|
| Level range | 0 Vpk to ±8 Vpk into 600 Ω |
| THD+N |
<-75 dB
for frequency 1 kHz and level 1 Vrms
|
| Traces / markers | 1 / 6 |
|---|---|
| Horizontal sweep per div |
1 µs to 100 ms/div
Accuracy <2%
|
| Vertical range |
1 mV/div to 20 V/div
Accuracy <5%
|
| Bandwidth | 20 kHz |
| Input range | 20 mV to 30 Vrms (42.4 Vpk) |
| Coupling | AC, DC |
| Trigger modes | Single, Normal, Automatic, Free run |
| Modes | P25, P25 Phase 2, DMR, NXDN, TETRA |
|---|---|
| P25 modulation fidelity accuracy | <5% of reading (2.5% to 12%) |
| P25 symbol deviation accuracy | ±1% |
| DMR FSK error range |
0 to 20%
Resolution 0.01%; accuracy <5% of reading (2.5 to 10%)
|
| DMR symbol deviation range |
1500 Hz to 2350 Hz
Accuracy ±10 Hz (1745 to 2140 Hz)
|
| DMR frequency error range |
±4000 Hz
Resolution 0.01 Hz
|
| DMR UUT TX/RX bit error rate range |
0 to 20%
Resolution 0.1%
|
| DMR protocol decode | Color Code, Call ID, Unit ID |
| Frequency range |
1 MHz to 6 GHz
Resolution 0.1 Hz
|
|---|---|
| Test port power, Port 1 | +10 dBm |
| Dynamic range | 90 dB |
| Parameters | S11, S21 |
| Measurements | Magnitude, VSWR, Distance to Fault, Cable Loss, Insertion Loss, Group Delay, Phase, S-Parameters Real and Imaginary |
| Calibration method | Short-Open-Load, Thru |
| Maximum distance | 1000 ft (305 m) |
| Weight | 15 lbs (6.8 kg) |
|---|---|
| Temperature, operating | 0°C to 50°C |
| Temperature, not operating |
-40°C to +71°C
Battery must not be subjected to temperatures below -20°C, nor above +60°C
|
| Relative humidity | 95% RH (non-condensing) |
| Altitude | 4600 m |
| Vibration, shock, bench handling, transit drop | MIL-PRF-28800F Class 3 |
| Type | Lithium Ion, 14.4 V, 6.8 Ah |
|---|---|
| Operating time | 2.3 hours typical with 2 batteries |
| Battery charging limits | 0°C to 45°C (32°F to 113°F) ≤85% RH |
| EMC | EMC IEC/EN 61326-1:2013, CISPR11:2009 +A1:2010 |
|---|---|
| Safety | EN 61010-1, 3rd Edition |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- LMR radio test and alignment
- PMR radio test and alignment
- P25 radio testing
- DMR radio testing
- NXDN radio testing
- TETRA radio testing
- Duplexer test and alignment
- Cable and antenna testing
Options & Accessories
| Option | Description |
|---|---|
| CX300-F6GHz | 3 GHz to 6 GHz frequency extension — Extends RF generator, receiver, power meter, RF error meter and spectrum analyzer frequency range to 3 GHz to 6 GHz |