Luna LWA 7600 Series Lightwave Analyzer (LWA 7000 Family) for C, C/L and O Band OFDR Component Test
OFDR-based lightwave analyzer series measuring distributed return loss, insertion loss and length of optical components in C, C/L and O bands.
- Wavelength range
- 1515 nm – 1570 nm
- Wavelength range
- 1520 nm – 1630 nm
- Wavelength range
- 1260 nm – 1360 nm
- Length measurement modes (in reflection)
- 20 m, 50 m, 100 m; Optional: 200 m, 500 m
Overview
The Luna LWA 7600 Series (LWA 7000 Family) is a lightwave analyzer series for testing passive optical components, modules and networks, including photonic integrated circuits (PICs) and silicon photonics devices. Models cover the C-band (LWA 7601-C), C/L-band (LWA 7601-CL) and O-band (LWA 7601-O) for telecom, datacom and silicon photonics applications.
Measurement principle
The instruments use optical frequency domain reflectometry (OFDR) to measure backscattered or transmitted light as a function of distance/time or wavelength. They measure and analyze the insertion loss (IL) and return loss (RL) distribution, frequency response and length of optical components scanned in either reflection or transmission mode. The series builds on the OFDR technology of Luna's Optical Backscatter Reflectometer (OBR) platforms.
Sampling resolution is 20 μm on the LWA 7601-C and 10 μm in fiber on the LWA 7601-CL and LWA 7601-O. Extended range options increase the maximum measurement length in reflection to 500 m (LWA 7601-C) or 1000 m (LWA 7601-CL and LWA 7601-O), which allows testing of fiber networks as well as components. Measuring RL, IL and length with a single instrument reduces test cost and complexity in production.
Software and connectivity
The instrument is operated from a controller (workstation-class laptop or rack-mountable controller) running the Lightwave Analyzer application software, with remote user interfaces available for Windows and Linux. Remote automation is provided through a SCPI API over TCP/IP.
Models in this series
| Model | Description | |
|---|---|---|
| LWA 7601-C |
Luna LWA 7601-C C-Band Lightwave Analyzer
C-band OFDR lightwave analyzer that measures distributed return loss, insertion loss and length of optical components with 20 μm sampling resolution. |
Details → |
| LWA 7601-C-RM |
Luna LWA 7601-C-RM C-Band Lightwave Analyzer, Rack-Mountable Mainframe
Rack-mountable C-band OFDR lightwave analyzer that measures distributed return loss, insertion loss and length with 20 µm sampling resolution. |
Details → |
| LWA 7601-CL-RM |
Luna LWA 7601-CL-RM Rack-Mountable C/L Band Lightwave Analyzer
Rack-mountable C/L band OFDR lightwave analyzer measuring distributed return loss, insertion loss and length of optical components and networks. |
Details → |
| LWA 7601-O |
Luna LWA 7601-O O-Band Lightwave Analyzer (LWA 7600 Series OFDR)
O-band OFDR lightwave analyzer that measures distributed return loss, insertion loss and length of optical components and networks. |
Details → |
| LWA 7601-CL |
LWA 7600 Series LWA 7601-CL C/L-Band Lightwave Analyzer
C/L-band OFDR lightwave analyzer that measures distributed return loss, insertion loss and length in reflection or transmission. |
Details → |
| LWA 7601-O-RM |
LWA 7600 Series LWA 7601-O-RM O-Band Rack-Mountable Lightwave Analyzer
Rack-mountable O-band OFDR lightwave analyzer that measures distributed return loss, insertion loss and length of optical components. |
Details → |
Key Features
- Return loss (RL) and insertion loss (IL) analysis
- Analyze components in reflection and transmission
- Trace distributed RL over length of optical path
- Spectral analysis of RL and IL
- Detect and precisely locate reflective events and measure path length
- Speed, resolution and accuracy for optimizing production test
- 20 μm sampling resolution on LWA 7601-C
- 10 μm sampling resolution in fiber on LWA 7601-CL and LWA 7601-O (4.29 μm in silicon photonics, ng = 3.5)
- 12 Hz scan/acquisition rate on LWA 7601-C
- 1 Hz scan/acquisition rate on LWA 7601-CL and LWA 7601-O
- Optional 19-inch rack-mount version for lab and production environments
- Reflection measurements: total IL, total path length, RL versus length, event loss (RL, IL), RL spectral amplitude analysis, event length
- Transmission measurements: spectral amplitude response
- Internal NIST-traceable HCN gas cell for wavelength and time-of-flight accuracy
- SCPI API over TCP/IP remote interface
Specifications
|
Wavelength range
LWA 7601-C |
1515 nm – 1570 nm |
|---|---|
|
Wavelength range
LWA 7601-CL |
1520 nm – 1630 nm |
|
Wavelength range
LWA 7601-O |
1260 nm – 1360 nm |
|
Length measurement modes (in reflection)
LWA 7601-C |
20 m, 50 m, 100 m; Optional: 200 m, 500 m |
|
Length measurement modes (in reflection)
LWA 7601-CL, LWA 7601-O |
20 m, 50 m, 100 m; Optional: 200 m, 500 m, 1000m |
|
Maximum measurement length, reflection
LWA 7601-C |
100 m / 500 m
500 m with extended range option OPT07650. Web page lists 1000 m; datasheet value used |
|
Maximum measurement length, reflection
LWA 7601-CL, LWA 7601-O |
100 / 200 / 500 / 1000 m
500 m / 1000 m with extended range option OPT07650 / OPT07651 |
|
Maximum measurement length, transmission
LWA 7601-C |
200 m / 1000 m
1000 m with extended range option OPT07650. Web page lists 2000 m; datasheet value used |
|
Maximum measurement length, transmission
LWA 7601-CL, LWA 7601-O |
200 / 400 / 1000 / 2000 m
Per length mode; longest modes with extended range option |
|
Sampling resolution (two-point)
LWA 7601-C |
20 µm / 80 µm
Distance between two sample points along the length axis in SMF-28; 80 µm with extended range option OPT07650 |
|
Sampling resolution (two-point)
LWA 7601-CL, LWA 7601-O |
10 / 20 / 80 / 80 µm
Distance between two sample points along the length axis in SMF-28; per length mode |
|
Wavelength accuracy
LWA 7601-C |
± 6 pm
Guaranteed via internal NIST-traceable HCN gas cell |
|
Wavelength accuracy
LWA 7601-CL, LWA 7601-O |
±1.5 pm
Guaranteed via internal NIST-traceable HCN gas cell |
|
Wavelength resolution (max)
LWA 7601-CL, LWA 7601-O |
0.02 pm |
|
Time-of-flight delay accuracy
LWA 7601-C |
± 0.001 %
Guaranteed via internal NIST-traceable HCN gas cell |
|
Time-of-flight delay accuracy
LWA 7601-CL, LWA 7601-O |
± 0.001 / ± 0.001 / ± 0.005 / ± 0.005 %
Per length mode; guaranteed via internal NIST-traceable HCN gas cell |
|
Center wavelength
LWA 7601-C |
1546.69 nm |
|
Center wavelength
LWA 7601-CL |
1566.35 nm |
|
Center wavelength
LWA 7601-O |
1305 nm |
|
Wavelength scan range
LWA 7601-C |
± 20 / ± 5
Standard / extended range option; unit not printed in datasheet |
|
Wavelength scan range
LWA 7601-CL |
± 40 / ± 20 / ± 5 / ± 5 nm
Per length mode |
|
Wavelength scan range
LWA 7601-O |
± 30 / ± 15 / ± 3.75 / ± 3.75 nm
Per length mode |
|
Measurement time (in continuous mode)
LWA 7601-C |
0.08 s (20 m), 0.4 s (100 m); 0.5 s
0.5 s with extended range option OPT07650 |
|
Measurement time (in continuous mode)
LWA 7601-CL, LWA 7601-O |
~1 (50 m), ~2.1 (100 m) / ~2.3 / ~1.3 / ~2.3 s
Per length mode |
| Maximum optical power | 5 mW |
|
RL dynamic range
LWA 7601-C |
70 dB
Range between strongest reflection greater than -60 dB and noise floor |
|---|---|
|
RL dynamic range
LWA 7601-CL, LWA 7601-O |
90 dB
Range between strongest reflection greater than -60 dB and noise floor |
|
Total range
LWA 7601-C |
0 to -130 dB
Noise floor return loss at half of maximum length |
|
Total range
LWA 7601-CL, LWA 7601-O |
0 to -140 dB
Noise floor return loss at half of maximum length |
|
Sensitivity
LWA 7601-C |
-135 dB
Noise floor return loss at half of maximum length |
|
Sensitivity
LWA 7601-CL, LWA 7601-O |
-145 dB
Noise floor return loss at half of maximum length |
|
Resolution
LWA 7601-C |
± 0.1 dB
Measured with 1 cm integration width |
|
Resolution
LWA 7601-CL, LWA 7601-O |
0.05 dB
Measured with 50 cm integration width, and 85 nm wavelength range / 10 THz optical frequency range |
|
Accuracy
LWA 7601-C |
± 0.5 dB
Measured with 1 cm integration width |
|
Accuracy
LWA 7601-CL, LWA 7601-O |
±1 dB
Measured with 50 cm integration width, and 85 nm wavelength range / 10 THz optical frequency range |
|
IL dynamic range, in transmission mode
LWA 7601-C |
70 dB |
|---|---|
|
IL dynamic range, in transmission mode
LWA 7601-CL, LWA 7601-O |
90 dB |
|
IL dynamic range, in reflection mode
LWA 7601-C |
15 dB
Two way loss before backscatter reaches noise floor and IL measurements are no longer possible |
|
IL dynamic range, in reflection mode
LWA 7601-CL, LWA 7601-O |
20 dB
Two way loss before backscatter reaches noise floor and IL measurements are no longer possible |
|
Resolution
LWA 7601-C |
± 0.1 dB
With 10 cm integration width for 20 m mode, and 40 cm for 500 m mode |
|
Resolution
LWA 7601-CL, LWA 7601-O |
0.05 dB
Localized event IL from fiber Rayleigh scatter RL before and after the event; sweep ≥10 THz, integration width ≥0.5 m |
|
Accuracy
LWA 7601-C |
± 0.2 dB
With 10 cm integration width for 20 m mode, and 40 cm for 500 m mode |
|
Accuracy
LWA 7601-CL, LWA 7601-O |
± 0.4 dB (typical ± 0.2)
Localized event IL from fiber Rayleigh scatter RL before and after the event; sweep ≥10 THz, integration width ≥0.5 m |
| Remote interface | SCPI API over TCP/IP |
|---|---|
| Optical connector type | FC/APC |
| Dimensions | 35 (L) x 47 (W) x 10 (H) cm |
|
Weight (controller not included)
LWA 7601-C |
15 lb (7 kg)
Printed as 15 (7) lb (kg) |
|
Weight (controller not included)
LWA 7601-CL, LWA 7601-O |
25 lb (11.4 kg)
Printed as 25 (11.4) lb (kg) |
Applications
- Spatial RL testing
- Automated IL test and analysis
- Skew measurement with sub-picosecond resolution
- PIC waveguide devices, PLCs, AWGs, ROADMs
- Silicon photonics
- Optical fiber networks, cable and subassemblies
- Telecom and datacom component manufacturing test
Options & Accessories
| Option | Description |
|---|---|
| OPT07650 | Extended range option — Extends measurement length of the LWA 7601-C to 500 m (in reflection); 500 m extended range for LWA 7601-CL and LWA 7601-O. |
| OPT07651 | Extended range option — Extends measurement length of the LWA 7601-CL and LWA 7601-O to 1000 m (in reflection). |