Field Master MS2070A Handheld RF Spectrum Analyzer, 9 kHz to 3 GHz
Handheld field-portable RF spectrum analyzer covering 9 kHz to 3 GHz with 32 GHz/s sweep speed and a 10.1-inch multi-touch screen.
- Frequency range
- 9 kHz to 3 GHz (Option 703)
- Tuning resolution
- 1 Hz
- Span
- 10 Hz to max frequency, Zero Span
- Frequency reference
- Internal, GNSS, External
Overview
The Anritsu Field Master MS2070A is a handheld, field-portable RF spectrum analyzer covering 9 kHz to 3 GHz. It is built on the same durable platform as the rest of the Field Master family and is intended for RF measurements in general-purpose field and lab environments, including routine spectrum monitoring and interference hunting in the crowded sub-3 GHz spectrum.
The analyzer captures and measures HF, VHF and UHF signals used in commercial and military applications. With the optional preamplifier, typical displayed average noise level is -167 dBm. Third-order intercept is +11 dBm typical at 2 GHz, and dynamic range is 105 dB typical.
The MS2070A performs standard swept spectrum analysis with sweep speeds up to 32 GHz/s. Smart measurements include channel power, occupied bandwidth, adjacent channel power, spectral emission mask, carrier-to-interference and field strength. Options add a high accuracy power meter, interference finder, channel scanner, GNSS receiver, coverage mapping and AM/FM modulation measurements.
Display and connectivity
The instrument uses a 10.1-inch capacitive multi-touch screen with 1280 x 800 resolution and supports several user-interface languages. Remote control is available over Ethernet and USBTMC using SCPI commands, with Wi-Fi as an option. PDF and HTML measurement reports can be generated directly on the instrument.
Key Features
- Spectrum analyzer from 9 kHz to 3 GHz
- Interference analyzer
- Smart measurements including Channel Power and Occupied Bandwidth
- True power meter with external USB power sensors
- AM/FM modulation quality measurements
- Coverage mapping
- Sweep speed of 32 GHz/s
- 20 MHz analysis bandwidth standard
- Dynamic range > 105 dB typical
- DANL -167 dBm typical with preamp on
- Resolution bandwidth 1 Hz up to 5 MHz
- Spectrogram display
- Trace recording and playback
- 10.1-inch high resolution multi-touch screen
- Multi-language support
- Ethernet and USBTMC remote control interfaces
- Optional Wi-Fi interface
- Built-in PDF/HTML report generation
- IP52 environmental protection in soft case
- USB 3.0 ports
Specifications
|
Frequency range
MS2070A-0703
|
9 kHz to 3 GHz (Option 703)
Usable to 0 Hz
|
|---|---|
| Tuning resolution | 1 Hz |
| Span | 10 Hz to max frequency, Zero Span |
| Frequency reference | Internal, GNSS, External |
| Internal frequency reference aging |
±1.0 x 10–6 per year
Standard TCXO
|
| Internal frequency reference accuracy |
±2.8 x 10–7 (-10 °C ± 55°C) plus aging
See GNSS Receiver (Option 31) for improved accuracy
|
| External frequency reference | 10 MHz, -10 dBm to +10 dBm |
|
Frequency accuracy with GNSS
MS2070A-0031
|
< ±2.5 x 10–8
With GNSS On, three minutes after satellite lock in selected mode (GNSS antenna connected)
|
| Analysis bandwidth | 20 MHz (standard) |
|---|---|
| Resolution bandwidth (RBW) | 1 Hz to 3 MHz, 1 Hz to 5 MHz in zero span |
| RBW selectivity | 4:1 nominal (–60 dB / –3 dB) |
| Video bandwidth (VBW) | 1 Hz to 3 MHz, 1 Hz to 5 MHz in zero span |
| CISPR bandwidth |
200 Hz, 9 kHz, and 120 kHz
Resolution bandwidth when using Quasi-Peak marker function
|
| VBW/Average type | Linear/Log |
| Sweep rate (non-zero span) | 32 GHz/s |
|---|---|
| Manual sweep | Maximum sweep time is 3600 s (1 hour) |
| Sweep points | 10 to 10,001 (1001 in zero span) |
| Zero span sweep time | 60 ns to 3600 s |
| Sweep time accuracy |
±2 %
In zero span
|
| SSB phase noise, 10 kHz offset |
–93 dBc/Hz maximum, –94 dBc/Hz typical
Offset from 1 GHz RF input
|
|---|---|
| SSB phase noise, 100 kHz offset |
–95 dBc/Hz maximum, –97 dBc/Hz typical
Offset from 1 GHz RF input
|
| SSB phase noise, 1 MHz offset |
–120 dBc/Hz maximum, –123 dBc/Hz typical
Offset from 1 GHz RF input
|
|
Residual spurious, preamp On
MS2070A-0008
|
< –105 dBm
RF input terminated, 0 dB input attenuation, > 10 MHz
|
| Residual spurious, preamp Off |
< –94 dBm
RF input terminated, 0 dB input attenuation, 10 MHz to 3 GHz
|
| Input-related spurious |
< –70 dBc
0 dB attenuation, –30 dBm input; exceptions listed in datasheet
|
| Local-oscillator related spurious | < –60 dBc nominal for offsets > 1 MHz |
| Dynamic range |
105 dB typical at 1 GHz
2/3 (TOI-DANL) in 1 Hz RBW; web page overview states >105 dB at 2.4 GHz
|
|---|---|
| Measurement range | DANL to +30 dBm |
| Display range | 1 to 15 dB/div in 1 dB steps, ten divisions displayed |
| Reference level range | –150dBm to +30dBm |
| Attenuator range | 0 to 50 dB, 5 dB steps |
| Reference level offset | 99.9 dB external loss to 99.9 dB external gain |
| Maximum continuous input | +30 dBm CW, ±50 VDC (≥ 10 dB attenuation); +23 dBm CW, ±50 VDC (< 10 dB attenuation); +10 dBm CW, ±50 VDC (preamp On) |
| Damage level | 5 W (+37 dBm) |
| Amplitude accuracy, 20 °C to 30 °C |
±1.00dB maximum, ±0.5dB typical
9 kHz to 3 GHz, after 30 minute warm-up; ≥ 10 dB attenuation, –50 dBm ≤ input ≤ –10 dBm, 1 kHz RBW; values below 100 kHz are with preamp off
|
| Amplitude accuracy, –10 °C to 55 °C |
±2.0dB maximum, ±0.5dB typical
9 kHz to 3 GHz, after 60 minute warm-up
|
|
DANL, preamp On, 100 kHz to 10 MHz
MS2070A-0008
|
–161 dBm maximum, –167 dBm typical
RMS detection, normalized to 1 Hz RBW
|
|
DANL, preamp On, 10 MHz to 2 GHz
MS2070A-0008
|
–160 dBm maximum, –165 dBm typical
RMS detection, normalized to 1 Hz RBW
|
|
DANL, preamp On, >2 GHz to 3 GHz
MS2070A-0008
|
–139 dBm maximum, –142 dBm typical
RMS detection, normalized to 1 Hz RBW
|
| DANL, preamp Off |
-150 typ
As stated on product page
|
| Third-order intercept (TOI), 1 GHz |
+7 dBm, Typical
–20 dBm tones 100 kHz apart, 0 dB input attenuation, preamp Off
|
| Third-order intercept (TOI), 2 GHz |
+11 dBm, Typical
–20 dBm tones 100 kHz apart, 0 dB input attenuation, preamp Off
|
| Third-order intercept (TOI), 3 GHz |
+14 dBm, Typical
–20 dBm tones 100 kHz apart, 0 dB input attenuation, preamp Off
|
| Second harmonic distortion, 50 MHz |
–65 dBc maximum
0 dB input attenuation, –30 dBm input, preamp Off
|
| Second harmonic distortion, > 50 MHz to 1.5 GHz |
–70 dBc, typical
0 dB input attenuation, –30 dBm input, preamp Off
|
| VSWR, 9 kHz to 2.0 GHz |
1.5:1 typical
≥10 dB input attenuation
|
| VSWR, 2 GHz to 3.0 GHz |
1.8:1 typical
≥10 dB input attenuation
|
|
Preamplifier frequency range
MS2070A-0008
|
9 kHz to 3 GHz |
| RF In | Type N(f), 50Ω |
|---|---|
| GPS | SMA(f) |
| External power | 5.5 mm barrel connector, 14 to 16 VDC |
| Ethernet interface |
RJ45 connector for Ethernet 10/100/1000 Mbps
Connect to PC or LAN for remote access
|
| USB interface | Two USB 3 Type A (supports file transfer); One USB 3 Type C (USBTMC) |
| Headset jack | 3.5 mm headset jack |
| External reference in | 10 MHz, SMA(f), 50Ω, maximum input +10 dBm |
| External trigger in | SMA(f), 50Ω, TTL-compatible levels |
| DC bias voltage |
+1 V to +34 V, Resolution: 0.1 V
SMA(f)
|
| Display | 10.1-inches capacitive touchscreen, 1280 x 800 resolution |
|---|---|
| Screen strength | IK08 (protected against a five joule impact) |
| Instrument memory |
8 GB
Nominally 6.5 GB available to users
|
| Language | English, Spanish, Chinese-simplified, Japanese, French, Korean |
|
WLAN (Wi-Fi)
MS2070A-0005
|
2x2 MIMO, 802.11 a/b/g/n/ac |
| Report format | PDF and HTML |
| Programming language | Standard Commands for Programmable Instruments (SCPI) |
| Remote control interfaces | Ethernet, WLAN (Option 5), USBTMC (USB C port) |
| Type | Li-Ion |
|---|---|
| Internal battery operation | 3 hours operation, typical |
| Nominal capacity | 8940 mAh |
| Nominal energy | 97 Wh |
| External battery accessory |
Six hours operation, typical
With an accessory battery contained in soft carrying case
|
| Charging temperature limit | 0 °C to +45 °C, relative humidity ≤80 % |
| Environmental standard | MIL-PRF-28800F Class 2 |
|---|---|
| Operating temperature range | –10ºC to 55ºC |
| Storage temperature range | –51ºC to 71ºC |
| Maximum relative humidity | 95 % RH at 30ºC, non-condensing |
| Altitude | 4600 meters, operating and non-operating |
| Half sine shock | 30 g |
| Ingress protection rating | Complies with IP52 when installed in soft carrying case |
| Size | 290 mm x 212 mm x 96 mm (11.4 in x 8.3 in x 3.7 in) |
|---|---|
|
Weight
MS2070A-0703
|
3.8 kg (8.39 lb) |
| Warranty | Standard three-year warranty; One-year warranty on battery |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Broadcast transmitter analysis
- Interference hunting and spectrum clearing
- EMC pre-compliance testing
- PIM hunting on rooftop cell sites
- 75 Ω cable TV testing
- Satellite communications
- Spectrum monitoring
Included
- 2000-2071-R MS2070A Soft Case
- 2000-1371-R Ethernet Cable, 2 m
- Certificate of Calibration and Conformance
- 2000-1859-R USB Cable, USB 3.0 Type-A to Type-C, 1 m
- 633-83 Li-ion Battery, 97 Wh
- 806-442-R SMA(m) to BNC(m) cable, 1 m
- 2000-2156-R SMA(m) to BNC(f) Adapter (qty 3)
- 40-204-R AC/DC Power Adapter
Options & Accessories
| Option | Description |
|---|---|
| MS2070A-0703 | 9 kHz to 3 GHz Spectrum Analyzer — Required |
| MS2070A-0005 | Wi-Fi Connectivity |
| MS2070A-0006 | Remove Wi-Fi and Bluetooth |
| MS2070A-0007 | Secure Data — Stores setup and measurement data only to external USB memory |
| MS2070A-0008 | Preamplifier |
| MS2070A-0017 | Secure Communication — Creates a secure, encrypted tunnel when connected to a network; USBTMC does not work with this option |
| MS2070A-0019 | High Accuracy Power Meter — Requires USB power sensor, sold separately. Available as 90-day time-limited option |
| MS2070A-0024 | Interference Finder — Requires Option 8; Option 31 and directional antenna recommended. Available as 90-day time-limited option |
| MS2070A-0027 | Channel Scanner — Available as 90-day time-limited option |
| MS2070A-0031 | GNSS Receiver — Requires GNSS antenna, sold separately. Available as 90-day time-limited option |
| MS2070A-0400 | Enable Vision Monitor — Available as 90-day time-limited option |
| MS2070A-0407 | Enable Vision High-Speed Port Scanner — Available as 90-day time-limited option |
| MS2070A-0431 | Coverage Mapping (Channel Power and RSSI only) — Requires Option 8 and 31. Available as 90-day time-limited option (MS2070A-9431) |
| MS2070A-0509 | AM/FM Modulation Measurements — Available as 90-day time-limited option |
| MS2070A-0703-0097 | Accredited Calibration to ISO17025 and ANSI/NCSL Z540-1 |
| MS2070A-0703-0098 | Standard Calibration to ISO17025 and ANSI/NCSL Z540-1 |
| MS2070A-0703-0099 | Premium Calibration to ISO17025 and ANSI/NCSL Z540-1 plus test data |