Anritsu MA24218A Microwave Universal USB Power Sensor with N male connector and silicone protective covering
MA24218A

MA24218A Microwave Universal USB Power Sensor, 10 MHz to 18 GHz True-RMS

True-RMS USB power sensor measuring average power from 10 MHz to 18 GHz over an 80 dB dynamic range (−60 dBm to +20 dBm).

Frequency Range
10 MHz to 18 GHz
Datasheet

Overview

The Anritsu MA24218A is a Universal USB power sensor for average power measurements from 10 MHz to 18 GHz over 80 dB of dynamic range. It uses a patented triple-path architecture that provides True-RMS measurements, similar to thermal sensors, across the entire frequency and dynamic range. This allows average power measurements of CW, multi-tone and digitally modulated signals independent of modulation bandwidth.

Digital processing inside the sensor supports measurement speeds of more than 1,600 continuous readings per second and more than 11,000 buffered readings per second. A low-noise design removes the need to zero the sensor before measurement for signals above −45 dBm, and no 1 mW reference calibration is required.

Internal and external triggering supports individual slot power measurements of TDMA waveforms and burst power measurements of periodic and non-periodic signals. External trigger latching handles pulses as narrow as 20 ns.

Control and connectivity

The sensor connects to a host over USB 2.0 and is controlled from a PC with remote programming commands or with the PowerXpert software application, which is supplied free of charge. It is also compatible with most Anritsu RF and microwave handheld instruments; a high-accuracy power meter software option such as Option 19 may be required on the instrument.

Damage protection is rated at +30 dBm CW and +34 dBm peak for pulses shorter than 10 µs. A removable silicone protective covering adds field durability and is required for warranted operation. Calibrations are NIST traceable, with a recommended 12-month calibration interval.

Key Features

  • Frequency range: 10 MHz to 18 GHz
  • Power measurement range: −60 dBm to +20 dBm (80 dB dynamic range)
  • True RMS measurements, independent of modulation
  • Fast measurement speed: >1,600 readings/s continuous, >11,000 readings/s buffered
  • Damage protection: +30 dBm CW, +34 dBm peak < 10 µs
  • No zeroing required for signals > −45 dBm
  • Elimination of 1 mW reference calibration
  • Advanced trigger capabilities for time-dependent power measurements
  • External trigger latching for pulses as narrow as 20 ns
  • NIST traceable calibration
  • Easy to use with a PC or select Anritsu handheld instruments; no benchtop power meter needed
  • Removable silicone protective covering for additional field durability
  • Controlled with PowerXpert software or remote programming commands
  • Worldwide calibration and service centers

Specifications

Frequency
Frequency Range 10 MHz to 18 GHz
Power Measurement
Dynamic Range –60 dBm to +20 dBm
Measurement Range 1 +20 dBm to +4 dBm
approximate
Measurement Range 2 <+4 dBm to –16 dBm
approximate
Measurement Range 3 <–16 dBm to –60 dBm
approximate
Ranging Auto and fixed ranging available
Damage Levels at RF Port +30 dBm, ±20 V DC
+34 dBm peak < 10 μs pulse and 10 % duty cycle; minimum
Response
Signal Channel Rise Time 8 μs
characteristic
Sampling Rate 140 kS/s
VSWR
VSWR, max (≤150 MHz) 1.17:1
VSWR, max (>150 MHz to 2 GHz) 1.12:1
VSWR, max (>2 GHz to 12.4 GHz) 1.22:1
VSWR, max (>12.4 GHz to 18 GHz) 1.25:1
Trigger
Source Bus, Continuous, Internal, External
Bus trigger is not available in PowerXpert application
Arm Type (for Internal/External) Auto, Single, Multiple, Standby
Internal Trigger
Dynamic Range –35 dBm to +20 dBm
Level Accuracy ±0.5 dB
characteristic
Slope Positive or Negative
Delay Range –5 ms to +10 s
Delay Resolution 10 μs
Hysteresis 0 to 10 dB, with 0.1 dB resolution
Trigger Hold Off 0 to 10 sec, with 0.01 ms resolution
External Trigger
External Trigger Input MCX (female), 5.5 V maximum
Impedance 4 kΩ
nominal
Type TTL/CMOS
Slope Positive or Negative
Delay Range –5 ms to +10 s
Delay Resolution 10 μs
High Level Input Voltage 2.3 V min, 3.0 V max
Low Level Input Voltage 1.3 V min, 1.6 V max
Latency 7.1 μs max
Latency is defined as the time delay between the defined edge of the applied trigger and the sensor switching into the triggered state
Trigger Pulse Width 20 ns min
Trigger Repetition Period 7.1 μs min
Trigger Hold Off 0 s to 10 s with 0.01 ms resolution
Measurement Uncertainty – Average Power (dB), 0 °C to 50 °C
–60 to <–16 dBm ≤0.05 GHz: 0.14; >0.05 to 2 GHz: 0.14; >2 to 12.4 GHz: 0.14; >12.4 to 18 GHz: 0.17
Two sigma confidence level, CW measurement after zero operation. Includes calibration factor and linearity over temperature uncertainties, but not mismatch, zero set and drift, or noise
–16 to <+4 dBm ≤0.05 GHz: 0.14; >0.05 to 2 GHz: 0.14; >2 to 12.4 GHz: 0.13; >12.4 to 18 GHz: 0.13
Two sigma confidence level, CW measurement after zero operation
+4 to +20 dBm ≤0.05 GHz: 0.14; >0.05 to 2 GHz: 0.15; >2 to 12.4 GHz: 0.15; >12.4 to 18 GHz: 0.14
Two sigma confidence level, CW measurement after zero operation
Measurement Uncertainty – Average Power (dB), 20 °C to 30 °C
–60 to <–16 dBm ≤0.05 GHz: 0.13; >0.05 to 2 GHz: 0.12; >2 to 12.4 GHz: 0.14; >12.4 to 18 GHz: 0.14
Two sigma confidence level, CW measurement after zero operation
–16 to <+4 dBm ≤0.05 GHz: 0.11; >0.05 to 2 GHz: 0.10; >2 to 12.4 GHz: 0.13; >12.4 to 18 GHz: 0.11
Two sigma confidence level, CW measurement after zero operation
+4 to +20 dBm ≤0.05 GHz: 0.11; >0.05 to 2 GHz: 0.10; >2 to 12.4 GHz: 0.10; >12.4 to 18 GHz: 0.11
Two sigma confidence level, CW measurement after zero operation
Zero and Noise
Zero Set, –60 to <–16 dBm 3.32E-10 W (–64.78 dBm)
Three sigma confidence level. One hour warm-up followed by a Zero operation; 256 averages, 40 ms aperture, temperature within ±1 °C
Zero Set, –16 to <+4 dBm 3.87E-08 W (–44.12 dBm)
Three sigma confidence level
Zero Set, +4 to +20 dBm 1.07E-06 W (–29.70 dBm)
Three sigma confidence level
Zero Drift, –60 to <–16 dBm 3.44E-10 W (–64.64 dBm)
Two sigma value of the reported power over one hour
Zero Drift, –16 to <+4 dBm 4.29E-08 W (–43.67 dBm)
Two sigma value of the reported power over one hour
Zero Drift, +4 to +20 dBm 9.96E-07 W (–30.02 dBm)
Two sigma value of the reported power over one hour
Noise, –60 to <–16 dBm 1.23E-10 W
Two sigma noise at 10.2 seconds of integration time; noise uncertainty expressed with three sigma confidence level
Noise, –16 to <+4 dBm 1.01E-08 W
Two sigma noise at 10.2 seconds of integration time
Noise, +4 to +20 dBm 8.56E-07 W
Two sigma noise at 10.2 seconds of integration time
Effect of Digital Modulation, –60 to <–16 dBm –0.048 to 0.080 dB
Measurement error with reference to a CW signal of equal power and frequency between 20 °C to 30 °C in Normal mode and average power ≤+20 dBm
Effect of Digital Modulation, –16 to <+4 dBm –0.038 to 0.088 dB
Measurement error with reference to a CW signal of equal power and frequency between 20 °C to 30 °C
Effect of Digital Modulation, +4 to +20 dBm –0.055 to 0.067 dB
Measurement error with reference to a CW signal of equal power and frequency between 20 °C to 30 °C
System
Measurand Average power
Measurement Resolution 0.01 dB max via PowerXpert, 0.001 dB max via remote command
Offset Correction –100 dB to +150 dB
Available only through the PowerXpert application; no remote command in the sensor firmware
Averaging Auto, Manual
Averaging Type Moving, Repeat
Number of Averages (Manual) 1 to 65,536
Maximum in Continuous Average and Timeslot modes is 65,536; in Scope mode the maximum equals 16,777,216 divided by the number of data points
Auto Average Resolution 1 dB, 0.1 dB, 0.01 dB
Averaging resolution of 0.001 dB is not available with the PowerXpert application
Auto Average Source Timeslot Number: 1 to 128; Scope Data Point Number: 1 to 16,384
Continuous Average Mode
Duty Cycle Correction 0.01 % to 100 %
Aperture Time 0.01 ms to 1 s
Measurement Speed, Continuous >1,600 readings/s
Minimum aperture, one average. Speed may vary depending on the speed of and load on the CPU controlling the sensor
Measurement Speed, Buffered >11,000 readings/s
Minimum aperture, one average
Buffer Size 8192
Scope Mode
Capture Time 0.01 ms to 1 s
Data Points 1 to 16,384
Resolution 0.01 ms max
Timeslot Mode
Maximum Number of Slots 128
Slot Width 0.01 ms to 100 ms
Maximum Capture Time 1000 ms
slot width x number of slots
Resolution 0.01 ms max via remote command; 0.01 ms max via PowerXpert
Exclusion Periods Start Exclusion: 0 ms to 10 ms; End Exclusion: 0 ms to 10 ms
List Mode
Number of Measurements 1 to 1000
Input Parameters Frequency (GHz), aperture time (ms), averages
PowerXpert PC Requirements
Processor and RAM Minimum: Equivalent to Intel Pentium III with 1 GB RAM or Intel Pentium IV with 512 MB RAM; Recommended: Equivalent to Intel Pentium IV with 1 GB RAM
PowerXpert version 3.0 or greater
Operating System Microsoft Windows 8, Windows 7, and Windows XP
Hard-Disk Free Space 100 MB minimum
Display Resolution 1024 × 768 minimum
Interface USB 2.0 high speed
General
RF Connector N male
Interface to Host USB 2.0 high speed
USB Micro-B port for connectivity to host (PC or Anritsu handheld instrument)
Current Consumption 410 mA to 450 mA
characteristic (20 ºC to 30 ºC)
Size 110 mm x 46 mm x 25.6 mm
excluding N connector and silicone protective covering
Weight 397 g (0.88 lb)
Warranty 1 year
Calibration Cycle 12 months
Anritsu recommended calibration interval
Warm-Up Time 60 minutes
Environmental
Operating Temperature Range 0 ºC to 50 ºC
Tests were performed per MIL-PRF-28800F (Class 3)
Storage Temperature Range –40 ºC to +71 ºC
Relative Humidity (non-condensing) 45 % at 50 ºC; 75 % at 40 ºC; 95 % at 30 ºC
Altitude 4600 m
operational max
Shock 30 g half-sine, 11 ms duration
Vibration Sinusoidal: 5 Hz to 55 Hz, 3 g max; Random: 10 Hz to 500 Hz, 2.34 g rms; Power Spectral Density: 0.01 g²/Hz
Regulatory Compliance
European Union EMC 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11
Low Voltage Directive 2014/35/EU
Safety EN 61010-1:2010
RoHS Directive 2011/65/EU
Australia and New Zealand RCM AS/NZS 4417:2012
South Korea KCC-REM-A21-0004

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • General purpose RF and microwave power measurement
  • Aerospace and defense
  • Satellite communications
  • Wireless communications
  • Time-dependent power measurements (GSM, WiMAX, TD-SCDMA, LTE)
  • TDMA slot power measurement
  • Burst power measurement of periodic and non-periodic waveforms
  • CW, multi-tone and digitally modulated signal measurement
  • Field measurements with Anritsu handheld instruments

Included

  • 10585-00021 Quick Start Guide
  • 2000-1605-R 1.5 m BNC(m) to MCX(m) cable
  • 2000-2010-R 1.83 m USB A to Micro-B cable

Options & Accessories

Option Description
MA24218A-097 Option 97: ISO/IEC 17025 and ANSI/NCSL Z540-1 or ANSI/NCSLI Z540.3 calibration — Includes test report, uncertainty data, and accreditation symbol
MA24218A-098 Option 98: Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
MA24218A-099 Option 99: Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 — Includes test report and uncertainty data
01-200 Calibrated torque wrench for N connector
01-204 Calibrated torque wrench for K and V connectors
3-1010-123 Power attenuator, DC to 8.5 GHz, 30 dB, 50 W, 50 Ω, N(m) to N(f)
3-1010-124 Power attenuator, DC to 8.5 GHz, 40 dB, 100 W, 50 Ω, N(m) to N(f)
3-1010-122 Power attenuator, DC to 12.4 GHz, 20 dB, 5 W, 50 Ω, N(m) to N(f)
42N50-20 Power attenuator, DC to 18 GHz, 20 dB, 5 W, 50 Ω, N(m) to N(f)
42N50A-30 Power attenuator, DC to 18 GHz, 30 dB, 50 W, 50 Ω, N(m) to N(f)
41KB-3 Power attenuator, DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f)
41KB-6 Power attenuator, DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f)
41KB-10 Power attenuator, DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f)
41KB-20 Power attenuator, DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f)
43KB-3 Power attenuator, DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f)
43KB-6 Power attenuator, DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f)
43KB-10 Power attenuator, DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f)
43KB-20 Power attenuator, DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f)
510-90-R Precision coaxial adapter, DC to 3.3 GHz, N(m) to 7/16 DIN(f)
510-91-R Precision coaxial adapter, DC to 3.3 GHz, N(f) to 7/16 DIN(f)
510-92-R Precision coaxial adapter, DC to 3.3 GHz, N(m) to 7/16 DIN(m)
510-93-R Precision coaxial adapter, DC to 3.3 GHz, N(f) to 7/16 DIN(m)
33NFNF50B Precision coaxial adapter, DC to 18 GHz, N(f) to N(f)
33NNF50B Precision coaxial adapter, DC to 18 GHz, N(m) to N(f)
33NN50B Precision coaxial adapter, DC to 18 GHz, N(m) to N(m)
34AN50 Precision coaxial adapter, DC to 18 GHz, GPC-7 to N(m)
34ANF50 Precision coaxial adapter, DC to 18 GHz, GPC-7 to N(f)
34NFK50 Precision coaxial adapter, DC to 18 GHz, N(f) to K(m)
34NFKF50 Precision coaxial adapter, DC to 18 GHz, N(m) to K(f)
34NK50 Precision coaxial adapter, DC to 18 GHz, N(m) to K(m)
34NKF50 Precision coaxial adapter, DC to 18 GHz, N(m) to K(f)
1091-26-R Precision coaxial adapter, DC to 18 GHz, N(m) to SMA(m)
1091-27-R Precision coaxial adapter, DC to 18 GHz, N(m) to SMA(f)
1091-80-R Precision coaxial adapter, DC to 18 GHz, N(f) to SMA(m)
1091-81-R Precision coaxial adapter, DC to 18 GHz, N(f) to SMA(f)

Documents & Resources