Anritsu Spectrum Master MS276xA ultraportable spectrum analyzer
Anritsu / MS2760A / Handheld
MS2762A-0090

Spectrum Master MS2762A-0090 Ultraportable Spectrum Analyzer, 6 GHz to 90 GHz

USB-powered ultraportable spectrum analyzer covering 6 GHz to 90 GHz with a W1 (1.0 mm) test port and –141 dBm typical DANL.

Frequency range
6 GHz to 90 GHz
Tuning resolution
1 Hz
Internal 10 MHz frequency reference
Accuracy: ±0.2 ppm (25 °C ± 25 °C) + aging; Aging: ±1.0 ppm/year
Frequency span
10 Hz to maximum frequency range of instrument
Datasheet

Overview

The Spectrum Master MS2762A-0090 is the 6 GHz to 90 GHz model of the Anritsu MS2762A ultraportable spectrum analyzer line, part of the MS276xA family. MS2762A models offer higher dynamic range and lower displayed average noise level than the MS2760A models and are intended for applications with demanding sensitivity requirements.

The analyzer uses Anritsu's patented nonlinear transmission line (NLTL) technology, also used in Anritsu vector network analyzers. Its pocket-sized form factor allows a direct connection to the device under test, which avoids long test port cables that add loss and error at millimeter-wave frequencies. The MS2762A-0090 has a W1 (1.0 mm) male test port connector.

The instrument is USB-powered and controlled from a Windows-based PC, laptop or tablet, so it can be used in the lab, on the manufacturing floor or in the field. Built-in measurements include channel power, adjacent channel power and occupied bandwidth, with spectrum and spectrogram displays, up to six traces and 12 markers.

Typical uses include 5G and millimeter-wave device development, automotive radar testing, on-wafer spectrum measurements together with Anritsu VNAs, 802.11ad/WiGig spectral mask testing, satellite communications, wireless backhaul and radio astronomy.

Key Features

  • Frequency coverage from 6 GHz to 90 GHz
  • W1 (1.0 mm) male test port connector
  • –141 dBm typical DANL at >70 GHz to 90 GHz (1 Hz RBW)
  • Dynamic range >108 dB typical at 70 GHz
  • Patented NLTL technology
  • Ultraportable form factor enables measurements right at the device under test
  • USB-powered and controlled from a Windows-based PC, laptop or tablet
  • Measure: channel power, adjacent channel power, occupied bandwidth
  • Spectrum and spectrogram displays
  • Up to 6 traces, 3 trace detectors and 12 markers
  • Resolution bandwidth 1 Hz to 3 MHz
  • External 10 MHz frequency reference input
  • External TTL trigger input

Specifications

Frequency
Frequency range 6 GHz to 90 GHz
Tuning resolution 1 Hz
Internal 10 MHz frequency reference Accuracy: ±0.2 ppm (25 °C ± 25 °C) + aging; Aging: ±1.0 ppm/year
Frequency span 10 Hz to maximum frequency range of instrument
Bandwidth
Resolution bandwidth (RBW) 1 Hz to 3 MHz (Span ≥10 Hz); Zero Span 5 kHz to 20 MHz
Video bandwidth (VBW) 1 Hz to 3 MHz (Span ≥10 Hz); Zero Span 5 kHz to 20 MHz
VBW/Average type Linear / Log
RBW filters Flat Top, Nuttall
Shape factor <5:1 typical
Amplitude
Dynamic range > 108 dB typical at 70 GHz
2/3 (TOI – DANL) in 1 Hz RBW
Display range 1 dB to 15 dB/div in 1 dB steps, ten divisions displayed
Measurement range DANL to 0 dBm
Reference level range –120 dBm to +30 dBm
Amplitude units dBm
Maximum safe input level +18 dBm CW, ±10 VDC
Amplitude accuracy, ≥ 6 GHz to 40 GHz ±1.8 dB maximum, ±0.5 dB typical (20 °C to 30 °C); ±3.0 dB maximum, ±1.0 dB typical (0 °C to 50 °C)
–10 dBm CW signal; 30 minute warm-up (20 °C to 30 °C), 60 minute warm-up (0 °C to 50 °C)
Amplitude accuracy, > 40 GHz to 70 GHz ±2.0 dB maximum, ±0.5 dB typical (20 °C to 30 °C); ±3.0 dB maximum, ±1.0 dB typical (0 °C to 50 °C)
–10 dBm CW signal
Amplitude accuracy, > 70 GHz to 90 GHz ±2.2 dB maximum, ±0.5 dB typical (20 °C to 30 °C); ±3.0 dB maximum, ±1.0 dB typical (0 °C to 50 °C)
–10 dBm CW signal
Displayed Average Noise Level (DANL)
DANL, > 6 GHz to 40 GHz –135 dBm maximum, –142 dBm typical
RMS detection, VBW/Avg type = Log, IF Gain On, 1 Hz RBW
DANL, > 40 GHz to 70 GHz –137 dBm maximum, –141 dBm typical
RMS detection, VBW/Avg type = Log, IF Gain On, 1 Hz RBW
DANL, > 70 GHz to 90 GHz –136 dBm maximum, –141 dBm typical
RMS detection, VBW/Avg type = Log, IF Gain On, 1 Hz RBW
Spectral Purity
SSB phase noise, 7 GHz 1 kHz: –80 max / –89 typ; 10 kHz: –93 max / –97 typ; 100 kHz: –94 max / –98 typ
20 °C to 30 °C
SSB phase noise, 30 GHz 1 kHz: –66 max / –74 typ; 10 kHz: –81 max / –88 typ; 100 kHz: –81 max / –88 typ
20 °C to 30 °C
SSB phase noise, 60 GHz 1 kHz: –60 max / –69 typ; 10 kHz: –75 max / –84 typ; 100 kHz: –75 max / –84 typ
20 °C to 30 °C
Third-order intercept (TOI), 18 GHz +22 dBm
IF Gain Off, typical, –10 dBm tones 1 MHz apart, –10 dBm reference level
Third-order intercept (TOI), 62 GHz +21 dBm
IF Gain Off, typical, –10 dBm tones 1 MHz apart, –10 dBm reference level
Second harmonic distortion –50 dBc maximum
At 7 GHz input, –25 dBm input
Residual spurs, 6 GHz to 70 GHz –85 dBm maximum, –95 dBm typical
Residual spurs, > 70 GHz to 90 GHz –84 dBm maximum, –95 dBm typical
Input-related spurious <–60 dBc
–20 dBm CW input; no image rejection is applied in zero span, so spurious impact may differ
Input Match (W1 Connector, typical)
6 GHz to 12.4 GHz VSWR 1.92:1, return loss 10 dB
> 12.4 GHz to 50 GHz VSWR 1.78:1, return loss 11 dB
> 50 GHz to 70 GHz VSWR 2.32:1, return loss 8 dB
> 70 GHz to 110 GHz VSWR 3.01:1, return loss 9 dB
Datasheet range; this model covers up to 90 GHz
Sweep and Trace
Zero span sweep time 0.02 ms to 60 s
Sweep (trace) points Settable range from 10 to 10,001 points; the default is 501 points
Minimum capture time 0 s to 10 s
Sweep time accuracy ± 2% in Zero Span
Traces Up to six traces
Detector type per trace Peak, RMS / Avg, Negative
Markers Up to 12 Markers
Smart measurements Channel Power, Occupied Bandwidth, Adjacent Channel Power
External Trigger
Source External, Free Run, Video, Periodic (0 s to 5 s)
Delay 0 ms to 1670 ms; –60 s (up to sweep time) in zero span and video or external trigger
Holdoff 0 ms to 5000 ms
Slope Rising, Falling, Both
Hysteresis 0 dB to 200 dB
Connectors
RF In W Connector (1.0 mm), male 50Ω
90 GHz and 110 GHz instruments; listed as W1* (m) on the product page
USB interface USB 3.0, Type C Connector
External reference in MCX(f), 50Ω, 10 MHz
External trigger in MCX(f), 50Ω, TTL Levels
Computer Requirement
Display resolution 16:9 / 16:10 Aspect Ratio (>1280 x 720 / 1280 x 800)
Operating system Windows® 7, 8.1, 10; 64-bit
Recommended minimum configuration Quad Core i7 fourth generation or higher CPU, 16 GB RAM, 128 GB Data Storage, USB 3.0
Environmental
Standard MIL-PRF-28800F Class 3
Not including Windows tablet/laptop/PC
Operating temperature range 0 ºC to 50 ºC
Storage temperature range –40 ºC to 71 ºC
Maximum relative humidity 95 % RH at 30 ºC, non-condensing
Vibration, sinusoidal 5 Hz to 55 Hz
Vibration, random 10 Hz to 500 Hz
Half sine shock 30 g
Altitude 4600 meters, operating and non-operating
Explosive atmosphere MIL-PRF-28800F Section 4.5.6.3; MIL-STD-810G, Method 511.5, Procedure 1
Regulatory Compliance
European Union EMC 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11; Low Voltage Directive 2014/35/EU; Safety EN 61010-1:2010; RoHS Directive 2011/65/EU
Not including Windows tablet/laptop/PC
Australia and New Zealand RCM AS/NZS 4417:2012
South Korea KCC-R-R-A2J-1015
Canada ICES-1(A)/NMB-1(A)
General
Size 155 mm x 84 mm x 27 mm (6.1 in x 3.3 in x 1.1 in)
Not including Windows tablet/laptop/PC
Weight 255 g (9.0 oz)
Warranty Standard one-year warranty
The product page states a standard three-year warranty; the technical data sheet states one year
Recommended calibration cycle 12 months

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • 5G and mmWave device development
  • Automotive radar testing
  • On-wafer spectrum measurements
  • 802.11ad/WiGig spectral mask testing
  • Satellite communications
  • Wireless backhaul
  • Radio astronomy
  • Antenna beam pattern testing
  • Microwave radio link testing
  • Electronic warfare and radar
  • RF component and device characterization on the production line

Included

  • USB 3.0 Type C to Type A Cable, 1 m (2000-1859-R)
  • BNC(m) to MCX(m) Cable, qty 2 (2000-1605-R)
  • Certificate of Calibration and Conformance

Options & Accessories

Option Description
MS2762A-0090-0097 Accredited Calibration to ISO17025 and ANSI/NCSL Z540-1 — Includes calibration certificate, test report, and uncertainty data
MS2762A-0090-0098 Standard Calibration to ISO17025 and ANSI/NCSL Z540-1 — Includes calibration certificate
MS2762A-0090-0099 Premium Calibration to ISO17025 and ANSI/NCSL Z540-1 — Includes calibration certificate, test report, and uncertainty data
34WV50 Precision Adapter, DC to 65 GHz, W1(m) to V(m), 50Ω
34WVF50 Precision Adapter, DC to 65 GHz, W1(m) to V(f), 50Ω
34WFV50 Precision Adapter, DC to 65 GHz, W1(f) to V(m), 50Ω
34WFVF50 Precision Adapter, DC to 65 GHz, W1(f) to V(f), 50Ω
33WW50 Adapter, W1(m) to W1(m)
33WWF50 Adapter, W1(m) to W1(f)
33WFWF50 Adapter, W1(f) to W1(f)
41W-3 Fixed Attenuator, DC to 110 GHz, 0.2 W, 3 dB, W1(m) to W1(f), 50Ω
41W-6 Fixed Attenuator, DC to 110 GHz, 0.2 W, 6 dB, W1(m) to W1(f), 50Ω
41W-10 Fixed Attenuator, DC to 110 GHz, 0.2 W, 10 dB, W1(m) to W1(f), 50Ω
W240A Precision Power Divider, DC to 110 GHz, W1(f) input, W1(f) outputs, 3 resistor, 50Ω
W241A Precision Power Splitter, DC to 110 GHz, W1(m) input, W1(f) outputs, 2 resistor, 50Ω
MN25110A Precision Directional Coupler, 20 GHz to 110 GHz, W1(f) input, W1(f) output, W1(f) coupled port, 50Ω
01-504 Torque Wrench, 0.45 N·m (4 lbf·in) for W1, 1.0 mm connectors
35WR10WF Precision Waveguide Coaxial Adapter, 75 GHz to 110 GHz, WR10 to W1(f)
SC7442 Precision Waveguide Coaxial Adapter, 60 GHz to 90 GHz, WR12 to W1(f)
1091-402-R Waveguide to Coaxial End Launch Adapter, 49.9 GHz to 75.8 GHz, WR15 to W1(f)
1091-401-R Waveguide to Coaxial End Launch Adapter, 60.5 GHz to 92.0 GHz, WR12 to W1(f)
1091-400-R Waveguide to Coaxial End Launch Adapter, 73.8 GHz to 110 GHz, WR10 to W1(f)
2000-1871-R Directional Horn Antenna, 49.9 GHz to 75.8 GHz, WR15, 25 dBi gain
2000-1872-R Directional Horn Antenna, 60.5 GHz to 92.0 GHz, WR12, 25 dBi gain
2000-1873-R Directional Horn Antenna, 73.8 GHz to 112.0 GHz, WR10, 25 dBi gain
3670W50-1 Test Port Cable, Armored, Semi-rigid, DC to 110 GHz, W1(m) to W1(f), 10 cm
3670W50-2 Test Port Cable, Armored, Semi-rigid, DC to 110 GHz, W1(m) to W1(f), 16 cm
3671W1-50-1 Test Port Cable, DC to 110 GHz, Flexible, W1(f) to W1(m), 10 cm
3671W1-50-2 Test Port Cable, DC to 110 GHz, Flexible, W1(f) to W1(m), 13 cm
3671W1-50-3 Test Port Cable, DC to 110 GHz, Flexible, W1(f) to W1(m), 16 cm
2000-1888-R USB 3.0 Powered Extension Cable, 10 m (32 ft) — Up to two can be used in series for a total length of 20 m
2000-1918-R 4-port 200 m USB Multimode Fiber Extender System

Documents & Resources

Spectrum Master Ultraportable Spectrum Analyzer MS276xA Product Brochure
brochure
Spectrum Master MS2762A Ultraportable mmWave Spectrum Analyzer Quick Fact Sheet
brochure
Spectrum Master MS276xA Technical Data Sheet
datasheet · PDF
Spectrum Master MS2760A Ultraportable mmWave Spectrum Analyzer Quick Fact Sheet
brochure · PDF
Antennas and Antenna Kits
datasheet · PDF
Low Noise Amplifiers Technical Data Sheet
datasheet · PDF
USB Extender Technical Data Sheet
datasheet · PDF
USB 3.0 Extension Cable 2000-1888-R Technical Data Sheet
datasheet · PDF
Ultraportable Spectrum Analyzer Declaration of Conformity
other · PDF
IoT, Medical & Healthcare Device Wireless Communication Quality Improvement
app note
Spectrum Monitoring with Field Master Pro Series Spectrum Analyzers
app note
Combining On-Wafer Vector Network Analyzer and Spectrum Analyzer Measurements to 110 GHz Application Note
app note
Automotive Radar Testing with Anritsu’s Spectrum Master MS2760A Ultraportable Spectrum Analyzer
app note
Millimeter Wave Measurement Solutions
brochure
Automotive Testing & Verification Solutions
brochure
Spectrum Master Ultraportable Spectrum Analyzer Product Flyer
brochure
Signal Analyzers and Spectrum Analyzers Selection Guide
brochure
Electronic Measuring Instruments Digest 2026 Catalog
brochure
Precision RF & Microwave Components Catalog
brochure
Automotive Test Solutions
brochure
Spectrum Master MS276xA User Guide
manual
Spectrum Master MS276xA Help
manual
Spectrum Master MS2760A Maintenance Manual
manual