Power Master MA24507A 9 kHz to 70 GHz mmWave Power Analyzer
USB-powered, frequency-selectable mmWave power analyzer covering 9 kHz to 70 GHz with a noise floor as low as –90 dBm.
- Frequency Range
- 9 kHz to 70 GHz
- RF Connector
- V(m) Connector (1.85 mm)
- Internal Reference Accuracy
- ±0.2 ppm
- Internal Reference Aging
- ±1.0 ppm/year aging
Overview
The Anritsu Power Master MA24507A is a USB-powered, frequency-selectable millimeter-wave power analyzer that covers 9 kHz to 70 GHz. It provides numeric, frequency-based average power measurements and can measure signals as low as –90 dBm. The input is a V(m) 1.85 mm connector.
The Power Master uses a receiver-based architecture rather than thermal sensing components. This shortens measurement times compared with thermal mmWave sensors and adds frequency selectivity, so power is measured only in the frequency band of interest. Up to six signals can be measured.
The instrument is slightly larger than a smartphone and takes up little bench space. It can be used for stand-alone measurements or as part of a system for calibrating vector network analyzers. Mounting holes allow it to be attached directly to probes for over-the-air or on-wafer testing.
PowerXpert software
The instrument is controlled with Anritsu's PowerXpert PC application. In addition to the general settings (measurement mode, center frequency, span and resolution), PowerXpert has a Power Hunter mode and a Channel Monitor mode. Power Hunter finds the six highest CW amplitudes and their frequencies within a user-defined range. Channel Monitor tracks the CW amplitude or channel power of up to six channels at once, each up to 20 MHz wide.
Key Features
- Measures signals as low as –90 dBm
- Suited to over-the-air testing of mmWave signals with high propagation loss
- User settings to control measurement speed and noise floor
- Channel Monitor mode in PowerXpert for monitoring up to six frequency channels at once
- Power Hunter mode in PowerXpert for searching up to six signals within a frequency range
- Mounting holes for direct mounting to probes for over-the-air or on-wafer testing
- Receiver-based architecture with frequency-selective measurement
- USB-powered, small form factor
Specifications
| Frequency Range | 9 kHz to 70 GHz |
|---|---|
| RF Connector | V(m) Connector (1.85 mm) |
| Internal Reference Accuracy |
±0.2 ppm
0 °C to 50 °C
|
| Internal Reference Aging | ±1.0 ppm/year aging |
| Continuous Mode Span (Channel Power Measurement) | 30 kHz to 2 GHz max |
| Continuous Mode Span (CW Max Measurement) | 10 kHz to Full Span |
| Channel Monitor Mode Span | 1 kHz to 20 MHz |
| Measurement Range | -90 — +10 dBm |
|---|---|
| Maximum Amplitude, ≤ 6.15 GHz |
+15 dBm
Characteristic
|
| Maximum Amplitude, > 6.15 GHz |
+10 dBm
Characteristic
|
| Damage Level (Continuous) | +30 dBm CW, +/- 10 VDC max |
| Range, Lower |
≤ –10 dBm
User-defined operating range; upper range recommended for signals above –10 dBm to avoid clipping or saturation
|
| Range, Upper | > –10 dBm |
| Average Noise Floor, Channel Power, 30 kHz span |
–88 dBm
Measured at 1 GHz center frequency
|
| Average Noise Floor, Channel Power, 10 MHz span |
–64 dBm
Measured at 1 GHz center frequency
|
| Average Noise Floor, Channel Power, 1 GHz span |
–40 dBm
Measured at 1 GHz center frequency
|
| Average Noise Floor, CW Max, High resolution |
–100 dBm
Measured at 1 GHz center frequency; 3 MHz span
|
| Average Noise Floor, CW Max, Medium resolution |
–90 dBm
Measured at 1 GHz center frequency; 3 MHz span
|
| Average Noise Floor, CW Max, Low resolution |
–80 dBm
Measured at 1 GHz center frequency; 3 MHz span
|
| Trigger Source | Bus, Continuous |
| 9 kHz to 12.4 GHz |
1.29:1 VSWR, 18 dB return loss
Typical
|
|---|---|
| > 12.4 GHz to 26.5 GHz |
1.43:1 VSWR, 15 dB return loss
Typical
|
| > 26.5 GHz to 40 GHz |
1.58:1 VSWR, 13 dB return loss
Typical
|
| > 40 GHz to 50 GHz |
1.67:1 VSWR, 12 dB return loss
Typical
|
| > 50 GHz to 70 GHz |
2.10:1 VSWR, 9 dB return loss
Typical
|
| Channel Power Measurement |
7 (300 kHz span), 20 (20 MHz span), 10 (1 GHz span)
Measured at 1 GHz center frequency; no averages
|
|---|---|
| CW Max Measurement (High) |
0.8 (300 kHz span), 15 (20 MHz span), 6 (1 GHz span)
Measured at 1 GHz center frequency; no averages
|
| CW Max Measurement (Medium) |
4 (300 kHz span), 25 (20 MHz span), 10 (1 GHz span)
Measured at 1 GHz center frequency; no averages
|
| CW Max Measurement (Low) |
20 (300 kHz span), 25 (20 MHz span), 10 (1 GHz span)
Measured at 1 GHz center frequency; no averages
|
| Amplitude Accuracy, 9 kHz to 644 MHz |
±1.3 dB max (±0.5 dB typ.) at 20 °C to 30 °C; ±2.0 dB max (±0.5 dB typ.) at 0 °C to 50 °C
After 30 / 60 minute warm-up respectively; excludes effects of noise and mismatch uncertainty
|
|---|---|
| Amplitude Accuracy, > 644 MHz to 40 GHz |
±1.8 dB max (±0.5 dB typ.) at 20 °C to 30 °C; ±3.0 dB max (±1.0 dB typ.) at 0 °C to 50 °C
After 30 / 60 minute warm-up respectively; excludes effects of noise and mismatch uncertainty
|
| Amplitude Accuracy, > 40 GHz to 70 GHz |
±2.0 dB max (±0.5 dB typ.) at 20 °C to 30 °C; ±3.0 dB max (±1.0 dB typ.) at 0 °C to 50 °C
Characteristic values between 67 and 70 GHz for MA24507A; excludes effects of noise and mismatch uncertainty
|
| Amplitude Accuracy (summary) |
±1.0 dB, typical
Web page summary; the datasheet gives the frequency-dependent values above
|
| Relative Power Accuracy, 9 kHz to < 6.15 GHz | ±0.3 dB |
| Relative Power Accuracy, 6.15 GHz to < 40 GHz | ±0.3 dB |
| Measurand | Channel power, CW peak power |
|---|---|
| Measurement Resolution | 0.01 dB max via PowerXpert, 0.01 dB max via remote command |
| Offset Correction |
–100 dB to +150 dB
Available only through the PowerXpert application; no remote command in the analyzer firmware
|
| Units | dBm, nW, µW, mW, W |
| Averaging | Manual, moving average, 1 to 1,000 averages |
| Continuous Mode Center Frequency | 9.5 kHz to (Max Freq – 500 Hz) |
| Continuous Mode Resolution | High, medium, low |
| Power Hunter Mode Measurement | CW max only |
| Power Hunter Start Frequency | 9 kHz to (Max Freq – 1 kHz) |
| Power Hunter Stop Frequency | 10 kHz to Max Freq |
| Power Hunter Set Minimum Power Range | –130 to 0 dBm |
| Channel Monitor Channel Frequencies | (9 kHz + Span/2) to (Max Freq – Span/2) |
| Channel Monitor Number of Channels | Up to 6 |
| Processor and RAM |
Equivalent to Quad Core i5 fourth generation or higher CPU, 8 GB RAM
PowerXpert version 4.0 or greater
|
|---|---|
| Operating System | Microsoft Windows 10, 8.1, or 7; 64-bit |
| Hard-Disk Free Space | 100 MB minimum |
| Display Resolution | 1024 × 768 minimum |
| Interface | USB 3.0 |
| RF Connector | V male (1.85 mm) |
|---|---|
| Trigger Port | Inactive |
| Reference Port | For Service Only |
| Interface to Host | USB 3.0 |
| Current Consumption | 900 mA max |
| Size | 155 mm x 84 mm x 27 mm (6.1 in x 3.3 in x 1.1 in) |
| Weight | 282 g (0.62 lb) |
| Warranty | 1 year |
| Warm-Up Time |
30 minutes
Specification condition
|
| Calibration Cycle |
12 months
Anritsu recommended calibration interval
|
| Operational Requirements | Tests were performed per MIL-PRF-28800F (Class 3). |
|---|---|
| Operating Temperature Range | 0 °C to 50 °C |
| Storage Temperature Range | –40 °C to +71 °C |
| Relative Humidity (non-condensing) | 45 % at 50 °C, 75 % at 40 °C, 95 % at 30 °C |
| Altitude | 4600 m operational max |
| Shock | 30 g half-sine, 11 ms duration |
| Vibration | Sinusoidal: 5 Hz to 55 Hz, 3 g max; Random: 10 Hz to 500 Hz, 2.34 g rms; Power Spectral Density: 0.01 g²/Hz |
| European Union EMC | 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11 |
|---|---|
| Low Voltage Directive | 2014/35/EU |
| Safety | EN 61010-1:2010 |
| RoHS Directive | 2011/65/EU |
| Australia and New Zealand | RCM AS/NZS 4417:2012 |
| South Korea | KCC-REM-A21-0004 |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Over-the-air mmWave testing
- On-wafer testing
- Vector network analyzer calibration systems
- Stand-alone RF and mmWave power measurement
- Antenna quality control for consumer electronics
Options & Accessories
| Option | Description |
|---|---|
| MA24507A-098 | Option 98: Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 |
| MA24507A-099 | Option 99: Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 — Includes test report and uncertainty data |
| 2000-1859-R | USB 3.0 Type C to Type A Cable, 1 m |
| 2000-1605-R | BNC(m) to MCX(m) Cable (qty 2) |
| 01-201 | Calibrated torque wrench for K and V connectors |
| 41VA-3 | Precision Fixed Attenuator, DC to 70 GHz, 1W, 3 dB, V(m) to V(f) |
| 41VA-6 | Precision Fixed Attenuator, DC to 70 GHz, 1W, 6 dB, V(m) to V(f) |
| 41VA-10 | Precision Fixed Attenuator, DC to 70 GHz, 1W, 10 dB, V(m) to V(f) |
| 41VA-20 | Precision Fixed Attenuator, DC to 70 GHz, 1W, 20 dB, V(m) to V(f) |
| 33VFVF50C | Precision Coaxial Adapter, DC to 70 GHz, V(f) to V(f), 50 Ω |
| 33VVF50C | Precision Coaxial Adapter, DC to 70 GHz, V(m) to V(f), 50 Ω |
| 35WR22VF | Waveguide to Coaxial Adapter (right angle), 33 GHz to 50 GHz, WR22 to V(f) |
| 35WR15VF | Waveguide to Coaxial Adapter (right angle), 50 GHz to 65 GHz, WR15 to V(f) |
| 35WR19VF | Waveguide to Coaxial Adapter (right angle), 40 GHz to 60 GHz, WR19 to V(f) |
| 1091-457-R | Waveguide to Coaxial End Launch Adapter, 39.3 GHz to 59.7 GHz, WR19 to V(f) |
| 1091-456-R | Waveguide to Coaxial End Launch Adapter, 49.9 GHz to 67.0 GHz, WR15 to V(f) |
| 1091-458-R | Waveguide to Coaxial End Launch Adapter, 33.0 GHz to 50.1 GHz, WR22 to V(f) |
| 1091-459-R | Waveguide to Coaxial End Launch Adapter, 26.4 GHz to 40.1 GHz, WR28 to V(f) |
| 2000-1867-R | Directional Horn Antenna, 17.6 GHz to 26.7 GHz, WR42, 25 dBi gain |
| 2000-1868-R | Directional Horn Antenna, 26.4 GHz to 40.1 GHz, WR28, 25 dBi gain |
| 2000-1869-R | Directional Horn Antenna, 33.0 GHz to 50.1 GHz, WR22, 25 dB gain |
| 2000-1870-R | Directional Horn Antenna, 39.3 GHz to 59.7 GHz, WR19, 25 dBi gain |
| 2000-1871-R | Directional Horn Antenna, 49.9 GHz to 75.8 GHz, WR15, 25 dBi gain |
| 2000-1888-R | USB 3.0 Powered Extension Cable, 10 m (32 ft) — Up to two can be used in series for a total length of 20 m |