VIAVI / Attenuators / MAP Series Modules
mVOA-C1

VIAVI MAP Series mVOA-C1 Variable Optical Attenuator Module

Stepper-motor, filter-based variable optical attenuator module for the VIAVI MAP chassis, in single-mode or multimode versions with 0.001 dB resolution.

Insertion loss at 0 dB
≤0.9 dB (≤1.5 dB tap option)
Insertion loss at 0 dB
≤1.5 dB
Insertion loss at 0 dB
≤1.5 dB
Insertion loss at 0 dB
≤2.4 dB
Datasheet

Overview

The VIAVI mVOA-C1 is a variable optical attenuator module for the Multiple Application Platform (MAP series). It uses a stepper motor and a filter to provide programmable attenuation for laboratory and manufacturing test. VIAVI describes the mVOA-C1 as the fifth generation of its attenuator line. Typical uses include amplifier testing, 100/400GE client optic testing and stress testing of coherent long-haul interfaces.

Measurement principle

The light passes through an expanded beam, and a linearly variable neutral density filter sets the attenuation. Motor control algorithms and a kinematic drive change the attenuation quickly with a short settling time. The module has two operating modes. In attenuation mode it changes the insertion loss without reference to the input power. In power mode it sets the absolute output power, which requires the internal power meter option. With input power tracking enabled, the module adjusts itself to keep the output power constant. Power and attenuation offsets can be set to compensate for losses in the test system.

Configurations

There are three configurations:

  • a standard attenuator
  • an attenuator with an output tap (5% for single-mode) for monitoring on an external power meter
  • an attenuator with built-in output power monitoring, which supports both attenuation mode and power mode

All three come in single-mode or multimode fiber and with several connector types. Each is available as a single, dual or quad unit, and every version fits in a single chassis slot. A single 8-slot MAP chassis can hold up to 48 independently controlled attenuators.

Platform

The mVOA belongs to the LightDirect module family. It is hot-pluggable and works with all current MAP-300 and MAP-200 chassis. MAP chassis are available in 2-, 3- and 8-slot versions. They are LXI compliant and provide SCPI-based automation drivers and a web-enabled multiuser interface. The graphical user interface has a summary view for working at system level and a detailed view for control of individual modules.

Key Features

  • Insertion loss <0.9 dB (single-mode standard configuration)
  • Flat spectral response for CWDM and DWDM multi-wavelength applications
  • Fast transition speed and settling time
  • Optional built-in power monitor for closed-loop power control
  • Optional high-power version handles up to 2 W input on single-mode fiber (500 mW on multimode)
  • Optional output tap for monitoring on an external power meter
  • Attenuation mode and output power mode with input power tracking
  • Multiple power and attenuation offsets to compensate for test system losses
  • Single, dual or quad configurations in one-slot modules
  • Up to 48 independently controlled attenuators in an 8-slot MAP chassis
  • Hot-pluggable, compatible with MAP-300 and MAP-200 chassis
  • Low back-reflection

Specifications

Optical
Insertion loss at 0 dB
Single-mode, Standard
≤0.9 dB (≤1.5 dB tap option)
Footnotes 1–4; excludes connectors, add 0.2 dB typically for connectors
Insertion loss at 0 dB
Single-mode, With Output Power Monitor
≤1.5 dB
Excludes connectors, add 0.2 dB typically for connectors
Insertion loss at 0 dB
Multimode, Standard
≤1.5 dB
Excludes connectors, add 0.2 dB typically for connectors
Insertion loss at 0 dB
Multimode, With Output Power Monitor
≤2.4 dB
Excludes connectors, add 0.2 dB typically for connectors
Polarization-dependent loss
Single-mode, Standard
≤ 0.08 dB (≤0.15 dB tap option)
For 0 to 25 dB
Polarization-dependent loss
Single-mode, With Output Power Monitor
≤0.15 dB
For 0 to 25 dB
Return loss
Single-mode
≥55 dB
Excludes connectors
Return loss
Multimode
≥45 dB
Excludes connectors
Maximum input power (standard power/high-power option)
Single-mode
+23 dBm/+33 dBm
Input to output port only
Maximum input power (standard power/high-power option)
Multimode
+23 dBm/+27 dBm
Input to output port only
Wavelength range
Single-mode
1260 to 1650 nm
Wavelength range
Multimode
750 to 1350 nm
Attenuation
Attenuation range
Single-mode
70 dB
Footnotes 1, 2
Attenuation range
Multimode
65 dB
Footnotes 1, 2
Shutter isolation
Single-mode
≥80 dB
Shutter isolation
Multimode
≥75 dB
Attenuation flatness
Single-mode
±0.04 dB
For range of 0 to 30 dB over 1480 to 1640 nm; for light with DOP <5%
Attenuation slew rate
Single-mode
≥25 dB/s
Attenuation slew rate
Multimode
≥20 dB/s
Relative attenuation uncertainty ±0.1 dB
For range of 0 to 45 dB, relative to 0 dB position, +23/–5°C only, DOP <5%, low-coherence laser source (>500 MHz). The datasheet introduction states an attenuation accuracy of ± 0.01 dB; the specification table value is used here.
Attenuation repeatability ±0.01 dB
For range of 0 to 45 dB, relative to 0 dB position, +23/–5°C only, low-coherence laser source (>500 MHz). The datasheet introduction states a repeatability of ±0.015 dB; the specification table value is used here.
Attenuation resolution 0.001 dB
0 to 65 dB for single-mode, 0 to 50 dB for multimode
Attenuation settling time ≤55 ms
Power Control (Output Power Monitor Option)
Closed-loop power range (standard power/high-power option)
Single-mode, With Output Power Monitor
+11 to -49 dBm / +31.5 to -28.5 dBm
Footnotes 1, 2
Closed-loop power range
Multimode, With Output Power Monitor
+5 to -40 dBm
Footnotes 1, 2
Power monitor linearity
With Output Power Monitor
±0.03 dB
+23/–5°C only; for light with DOP <5%
Power setting repeatability
With Output Power Monitor
±0.015 dB
For light with DOP <5%
Power setting resolution
With Output Power Monitor
0.001 dB
General
Warm-up time 30 min
Calibration period 3 Years
Chassis compatibility All current MAP-300 and MAP-200 chassis
Compliance CE, CSA/UL/IEC61010-1, plus LXI class C requirements when installed in a MAP chassis
Environmental
Operating temperature 0 to 50°C
Storage temperature -30 to 60°C
Operating humidity 15 to 80% RH, 0 to 40°C noncondensing
Physical
Dimensions (W x H x D) 4.1 x 13.3 x 37.0 cm
Weight 1.1 kg (single) / 1.3 kg (dual) / 1.7 kg (quad)

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • Optical amplifier (EDFA) testing
  • 100/400GE client optic testing
  • Stress testing of coherent long-haul interfaces
  • CWDM and DWDM multi-wavelength testing
  • Manufacturing test systems
  • Laboratory optical test

Options & Accessories

Option Description
MVOA-C1SS0-MXXX-MYY Single VOA, Standard Power, No Option — XXX = fiber type code, YY = connector type code
MVOA-C1DS0-MXXX-MYY Dual VOA, Standard Power, No Option — XXX = fiber type code, YY = connector type code
MVOA-C1QS0-MXXX-MYY Quad VOA, Standard Power, No Option — XXX = fiber type code, YY = connector type code
MVOA-C1SSM-MXXX-MYY Single VOA, Standard Power, Monitor Option — XXX = fiber type code, YY = connector type code
MVOA-C1DSM-MXXX-MYY Dual VOA, Standard Power, Monitor Option — XXX = fiber type code, YY = connector type code
MVOA-C1QSM-MXXX-MYY Quad VOA, Standard Power, Monitor Option — XXX = fiber type code, YY = connector type code
MVOA-C1SS1-M100-MYY Single VOA, Standard Power, Tap Option, Single Mode Fiber — YY = connector type code
MVOA-C1DS1-M100-MYY Dual VOA, Standard Power, Tap Option, Single Mode Fiber — YY = connector type code
MVOA-C1QS1-M100-MYY Quad VOA, Standard Power, Tap Option, Single Mode Fiber — YY = connector type code
MVOA-C1SH0-MXXX-MYY Single VOA, High Power, No Option — XXX = fiber type code, YY = connector type code
MVOA-C1DH0-MXXX-MYY Dual VOA, High Power, No Option — XXX = fiber type code, YY = connector type code
MVOA-C1SHM-MXXX-MYY Single VOA, High Power, Monitor Option — XXX = fiber type code, YY = connector type code
MVOA-C1DHM-MXXX-MYY Dual VOA, High Power, Monitor Option — XXX = fiber type code, YY = connector type code
MSUP-300A-4VOAPWC MAP-300 License to enable 4 VOA cassettes to use external power control
MSUP-300A-8VOAPWC MAP-300 License to enable 8 VOA cassettes to use external power control
M100 Fiber type: 9µm Single Mode — XXX code
M101 Fiber type: 50µm (OM3) — XXX code
M102 Fiber type: 62.5µm (OM1) — XXX code
MFP Connector type: FC/PC — YY code
MFA Connector type: FC/APC — YY code
MSC Connector type: SC/PC — YY code
MSU Connector type: SC/APC — YY code
MLC Connector type: LC/PC — YY code
MLU Connector type: LC/APC — YY code
AC500 FC/PC-FC/PC Universal Connector Adapter — Replacement mating sleeve
AC501 FC/PC-SC/PC Universal Connector Adapter — Replacement mating sleeve
AC502 FC/APC-FC/APC Universal Connector Adapter — Replacement mating sleeve
AC503 FC/APC-SC/APC Universal Connector Adapter — Replacement mating sleeve

Documents & Resources