PolaDetect POD-001 High Speed In-Line Polarimeter
In-line polarimeter that outputs four voltage signals for measuring degree of polarization (DOP) and state of polarization (SOP) in microseconds.
- Wavelength range
- 1550 ±50 nm standard
- Insertion loss (IL)
- 0.8 dB typical, 1.2 dB max.
- Return loss (RL)
- 55 dB
- PDL
- <0.25 dB
Overview
The Luna POD-001 is an in-line polarimeter designed for low-cost, high-speed polarization characterization without interrupting data traffic. It outputs four voltage signals that are used to calculate both the degree of polarization (DOP) and the state of polarization (SOP) of the light passing through the device, with results available in microseconds.
The device, marketed as PolaDetect™, is intended for integration into polarization monitoring and polarization stabilization modules, or into polarization characterization instruments. It is supplied with a preamplification board that provides analog signals for SOP/DOP calculation, feedback control and a computer interface. A calibration matrix for the calculation is provided with every device.
Versions without the preamplification board and calibration matrix are also available for OEM purposes.
The POD-001 has been discontinued by the manufacturer. The POD-201 polarimeter is listed as a similar product.
Key Features
- High speed and low loss
- Compact size
- No moving parts
- Four voltage outputs for DOP and SOP calculation
- Preamplification board for analog SOP/DOP signals, feedback control and computer interface
- Calibration matrix provided with every device
- OEM versions available without preamplification board and calibration matrix
Specifications
| Wavelength range | 1550 ±50 nm standard |
|---|---|
| Insertion loss (IL) | 0.8 dB typical, 1.2 dB max. |
| Return loss (RL) | 55 dB |
| PDL | <0.25 dB |
| PMD | <0.1 ps |
| Measurement bandwidth, optical head |
1.5 MHz
Optical head
|
|---|---|
| Measurement bandwidth, with preamp board |
50 kHz
With preamp board
|
| SOP uncertainty | 1% max |
| DOP uncertainty | 2% max |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Polarization-sensitive optical coherence tomography (OCT)
- OEM modules for fiber optic sensing
- Optical measurement systems
- Polarization monitoring modules
- Polarization stabilization modules
- Polarization characterization instruments
Included
- Preamplification board
- Calibration matrix