VectorStar MS4647B 70 GHz Microwave Vector Network Analyzer
2-port VectorStar vector network analyzer covering 10 MHz to 70 GHz, optionally from 70 kHz, with V (male) test ports.
- Frequency range
- 10 MHz to 70 GHz
- Test ports
- 2-port, V(male)
- Recommended calibration cycle
- 12 months
- Warm-up time
- 90 minutes
Overview
The Anritsu MS4647B is the 70 GHz model of the VectorStar MS4640B series of microwave vector network analyzers. It is a 2-port instrument with V (male) test ports covering 10 MHz to 70 GHz, and Option 070 extends the low end of the range to 70 kHz.
The VectorStar platform is built for S-parameter measurements of RF, microwave and millimeter-wave devices. Its users include device modeling engineers, R&D engineers and manufacturing test engineers who need accurate results without giving up throughput.
Measurement applications
Software options add pulsed measurements (PulseView), intermodulation distortion analysis with an internal combiner (IMDView), and differential device analysis with dual internal sources (DifferentialView). Further options add noise figure and differential noise figure, eye diagrams, time domain analysis and Universal Fixture Extraction for de-embedding test fixtures.
Spectrum analysis and expansion
Option 49 adds a spectrum analyzer for spectral-domain measurements such as harmonics, spurs and distortion products. The instrument can serve as the base of the ME7838 broadband systems and of multiport configurations built with the MN469xC test sets. When paired with the MN4765B O/E calibration module, it also supports optoelectronic measurements.
Key Features
- Frequency coverage 10 MHz to 70 GHz, optional 70 kHz low-end extension (Option 070)
- 2-port configuration with V (male) test ports
- PulseView pulse measurements with 2.5 ns pulse resolution and 100 dB dynamic range
- IMDView with internal combiner for switching between S-parameter and IMD measurements in a single connection
- DifferentialView real-time analysis of differential devices with the dual internal source option
- Spectrum analyzer option from 70 kHz to 70 GHz (MS4647B-049)
- Noise figure option based on a cold source technique
- Differential Noise Figure option
- Universal Fixture Extraction (UFX) de-embedding software option
- Internal trace-based eye diagram option
- Optoelectronic measurements using the MN4765B O/E Calibration Module
- 3-year warranty
Specifications
| Frequency range |
10 MHz to 70 GHz
Optional 70 kHz low end (Option 070)
|
|---|---|
| Test ports | 2-port, V(male) |
| Recommended calibration cycle | 12 months |
| Warm-up time |
90 minutes
Before specifications apply
|
| 0.07 to 0.3 MHz | 85 dB |
|---|---|
| > 0.3 to 2 MHz | 102 dB |
| > 2 to 10 MHz | 115 dB |
| > 0.01 to < 2.5 GHz | 122 dB |
| 2.5 to 5 GHz | 116 dB |
| > 5 to 20 GHz | 115 dB |
| > 20 to 38 GHz | 116 dB |
| > 38 to 50 GHz | 115 dB |
| > 50 to 65 GHz | 110 dB |
| > 65 to 67 GHz | 108 dB |
| > 67 to 70 GHz | 107 dB |
|
> 0.01 to < 2.5 GHz, at b1 or b2
Option 51
|
140 dB
Option 51
|
|
> 0.01 to < 2.5 GHz, at b1 or b2, with Option 31
Option 31, Option 51
|
142 dB
Option 51 with Option 31
|
| 0.07 to 0.3 MHz |
–75 dBm
10 Hz IF bandwidth, no averaging, –10 dBm port power, RMS
|
|---|---|
| > 0.3 to 2 MHz | –92 dBm |
| > 2 to 10 MHz | –105 dBm |
| > 0.01 to 70 GHz | –110 dBm |
|
> 0.01 to < 2.5 GHz, at bx loops
Option 51
|
–129 dBm
Option 51; characteristic
|
| 70 kHz to 0.01 GHz | +10 to –25 dBm |
|---|---|
| > 0.01 to < 2.5 GHz | +12 to –25 dBm |
| 2.5 to 5 GHz | +6 to –20 dBm |
| > 5 to 20 GHz | +5 to –20 dBm |
| > 20 to 38 GHz | +6 to –25 dBm |
| > 38 to 50 GHz |
+5 to –25 dBm
Rated power is typical 49 GHz to 50 GHz
|
| > 50 to 65 GHz | 0 to –25 dBm |
| > 65 to 67 GHz | –2 to –25 dBm |
| > 67 to 70 GHz | –3 to –25 dBm |
|
> 0.01 to < 2.5 GHz, with Option 31
Option 31
|
+14 to –25 dBm |
|
> 0.01 to < 2.5 GHz, Option 61 or 62
Option 61, Option 62
|
+10 to –85 dBm
Applies to Port 1
|
| Default power, standard |
–3 dBm
–5 dBm for MS4647B Option 8x systems
|
|---|---|
|
Default power, Option 51, 61, or 62
Option 51, Option 61, Option 62
|
–10 dBm |
| Power accuracy, 70 kHz to 0.01 GHz |
± 1.5 dB
Measured at default power
|
| Power accuracy, > 0.01 to 40 GHz | ± 1.5 dB |
| Power accuracy, > 40 to 67 GHz | ± 3.0 dB |
| Power accuracy, > 67 to 70 GHz |
± 4.0 (± 3.0) dB
Typical in parentheses
|
| Power resolution | 0.01 dB |
| Frequency resolution | 1 Hz |
|---|---|
| Frequency accuracy |
± 5 x 10-7 Hz/Hz
At time of calibration
|
| Frequency stability | < 5 x 10-9/°C over 0 °C to 50 °C temperature; < 1 x 10-9/day aging, instrument on |
| 0.07 to 0.3 MHz |
+5 dBm
0.3 dB compression for < 0.3 MHz; characteristic
|
|---|---|
| > 0.3 MHz to 70 GHz |
+10 dBm
Characteristic
|
| > 500 kHz to 20 GHz |
< 0.0045 dB, < 0.05 degree
1 kHz IF bandwidth, default power, RMS; characteristic on MS4647B
|
|---|---|
| > 20 to 40 GHz | < 0.006 dB, < 0.06 degree |
| > 40 to 67 GHz | < 0.006 dB, < 0.08 degree |
| > 67 to 70 GHz |
< 0.008 (< 0.006) dB, < 0.08 degree
Typical in parentheses
|
| Directivity, 70 kHz to 0.01 GHz | > 38 dB |
|---|---|
| Directivity, > 0.01 to 20 GHz | > 41 dB |
| Directivity, > 20 to 40 GHz | > 37 dB |
| Directivity, > 40 to 67 GHz | > 35 dB |
| Directivity, > 67 to 70 GHz | > 30 dB |
| Source match, > 40 to 67 GHz | > 28 dB |
| Source match, > 67 to 70 GHz | > 26 dB |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- S-parameter measurements of RF, microwave and millimeter-wave devices
- Device modeling
- Pulsed device measurements
- Intermodulation distortion measurements
- Differential device characterization
- Noise figure measurements
- Signal integrity and eye diagram analysis
- Optoelectronic (E/O and O/E) device measurements
- Material measurements
- Antenna measurements
- On-wafer measurements
Options & Accessories
| Option | Description |
|---|---|
| Option 070 | 70 kHz Low-end Frequency Extension |
| MS4647B-049 | 70 kHz to 70 GHz Spectrum Analyzer option |
| Option 001 | Rack Mount |
| Option 002 | Time Domain with advanced capabilities |
| Option 007 | Receiver Offset with Multiple Source Control for frequency-translated measurements |
| Option 021 | Universal Fixture Extraction |
| Option 031 | Dual Source |
| Option 032 | Internal Combiner |
| Option 036 | Extended Memory — Must be purchased with Option 35 |
| Option 041 | Noise Figure |
| Option 042 | PulseView |
| Option 043 | DifferentialView |
| Option 044 | IMDView |
| Option 046 | Fast CW |
| Option 047 | Eye Diagram |
| Option 048 | Differential Noise Figure |
| Option 053 | External ALC leveling |
| Option 05x | Direct Access Loops |
| Option 06x | Active Measurement Suites with 2 or 4 attenuators |
| Option 08x | Broadband/Millimeter Wave Interface |
| MN469xC | 4-port Test Sets, K and V — For 70 kHz to 70 GHz multi-port solutions |
| MN4765B | O/E Calibration Module — Options available for 40, 70 GHz and 110 GHz, at 850, 1060, 1310, and 1550 nm wavelengths |