Anritsu MA2491A Wideband Peak and Average Power Sensor, 50 MHz to 18 GHz
Wideband peak and average power sensor covering 50 MHz to 18 GHz, with 20 MHz video bandwidth, for ML2480 and ML2490 series power meters.
- Frequency Range
- 50 MHz - 18 GHz
- Video Bandwidth
- 20 MHz
Overview
The Anritsu MA2491A is a wideband power sensor in the MA249xA series. It is designed for use with the ML2480 and ML2490 series power meters and covers a frequency range of 50 MHz to 18 GHz.
The sensor measures peak power, crest factor, average power, rise time, fall time, maximum power and minimum power, and provides statistical data on wideband signals. This makes it suitable for characterizing radar and modulated communication signals.
When used with ML2480 and ML2490 series power meters, the sensor supports a sampling rate of 64 MS/s and a 20 MHz video bandwidth. It connects to the device under test through a precision N(m) RF connector with a hexagon coupling nut, so it can be attached with a standard torque wrench.
Key Features
- Peak, crest factor, average, rise time, fall time, maximum and minimum power measurements
- Statistical data of wideband signals
- Suitable for radar and communication signals such as WiMax, WCDMA, WLAN and GSM
- CW and average power measurements as low as -60 dBm
- 20 MHz video bandwidth
- Sampling rate of 64 MS/s with ML2480 and ML2490 series power meters
- Precision RF connector with hexagon coupling nut for torque wrench attachment
Specifications
| Frequency Range | 50 MHz - 18 GHz |
|---|---|
| Video Bandwidth | 20 MHz |
| Dynamic Range |
-60 to +20 dBm
CW Mode
|
|---|---|
| Sensor Linearity | <7% 50 to 300 MHz; <3.5% 0.3 to 18 GHz |
| SWR | <1.17; 50-150 MHz, <1.12; 0.15-2.5 GHz, <1.22; 2.5-12.4 GHz, <1.25; 12.4-18 GHz |
| Rise Time |
<18 ns
10 dBm, room temperature, ML2480 series power meter
|
|---|---|
| Sampling Rate |
64 MS/s
With ML2480 and ML2490 series power meters
|
| RF Connector |
N(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
|
|---|
| EMC Directive | 2014/30/EU |
|---|---|
| Low Voltage Directive | 2014/35/EU |
| RoHS Directive | 2011/65/EU & 2015/863 |
| Electromagnetic Compatibility | EN 61326-1:2013 |
| Emissions | EN 55011:2009 +A1:2010 Group 1 Class A |
| Immunity | EN 61000-4-2:2009 4 kV CD, 8 kV AD; EN 61000-4-3:2006 +A2:2010 3 V/m; EN 61000-4-4:2012 0.5 kV S-L, 1 kV P-L; EN 61000-4-5:2014 0.5 kV L-L, 1 kV L-E; EN 61000-4-6:2014 3 V; EN 61000-4-8:2010 3 A/m; EN 61000-4-11: 2004 100% @ 20 ms |
| Electrical Safety | EN 61010-1:2010 |
| Warranty | 1 Year Warranty |
|---|
Applications
- Radar signal measurement
- WiMax signal measurement
- WCDMA signal measurement
- WLAN signal measurement
- GSM signal measurement