Anritsu MA2482D Universal Power Sensor (Average), 10 MHz to 18 GHz
True-RMS universal average power sensor covering 10 MHz to 18 GHz with an 80 dB dynamic range from -60 to +20 dBm.
- Frequency Range
- 10 MHz - 18 GHz
- Dynamic Range
- -60 to +20 dBm
- Detection
- RMS
- SWR (10-150 MHz)
- <1.17
Overview
The Anritsu MA2482D is a member of the MA248xD series of universal power sensors for average power measurement. It covers 10 MHz to 18 GHz and offers a dynamic range of 80 dB, from -60 to +20 dBm. It can measure average power on CW, multi-tone and modulated RF waveforms such as 3G, 4G and OFDM signals.
The sensor architecture uses three pairs of diodes, each configured to operate in its square-law region over the sensor's dynamic range. As a result, the sensor measures true RMS power regardless of the type or bandwidth of the modulation on the input signal.
Option 1 (MA2480/01) adds a fast CW mode for high-speed measurement of CW signals, along with TDMA and pulse measurement capability. With this option, one of the diode pairs is calibrated for linearity over a wide dynamic range. Calibration factors are stored in the sensor's EEPROM.
Key Features
- True-RMS detection for accurate average power measurements of any signal type
- 80 dB dynamic range
- Frequency coverage from 10 MHz to 18 GHz
- Option 1 enables fast and accurate (1.8 % linearity) CW average power measurements
- Option 1 adds TDMA and pulse measurement capability
- Calibration factors stored in EEPROM
- Precision RF connector with hexagon coupling nut for use with an industry-standard torque wrench
Specifications
| Frequency Range | 10 MHz - 18 GHz |
|---|---|
| Dynamic Range | -60 to +20 dBm |
| Detection |
RMS
Listed as Video Bandwidth: RMS
|
| SWR (10-150 MHz) | <1.17 |
| SWR (0.15-2 GHz) | <1.12 |
| SWR (2-12.4 GHz) | <1.22 |
| SWR (12.4-18 GHz) | <1.25 |
|
Rise Time
MA2480/01
|
<0.004 ms
With option 1 only; 0.0 dBm, room temperature
|
| Sensor Linearity (10 MHz to 6 GHz) | 3% (-60 to +20 dBm) |
| Sensor Linearity (6 to 18 GHz) |
3% (-60 to +20 dBm); 3.5% (0 to +20 dBm)
Manufacturer page prints the second range as 'o to +20 dBm'
|
|
Sensor Linearity (CW, Option 1)
MA2480/01
|
1.8% CW
With option 1
|
| RF Connector |
N(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
|
|---|
| EMC Directive | 2014/30/EU |
|---|---|
| Low Voltage Directive | 2014/35/EU |
| RoHS Directive | 2011/65/EU & 2015/863 |
| Electromagnetic Compatibility | EN 61326-1:2013 |
| Emissions | EN 55011:2009 +A1:2010 Group 1 Class A |
| Electrical Safety | EN 61010-1:2010 |
| Warranty | 1 Year |
|---|
Applications
- Average power measurement of CW signals
- Multi-tone signal power measurement
- Modulated RF signal measurement (3G, 4G, OFDM)
- TDMA and pulse measurements (with Option 1)
Options & Accessories
| Option | Description |
|---|---|