Anritsu MA248xD series universal average power sensors, including the MA2482D
MA2482D

Anritsu MA2482D Universal Power Sensor (Average), 10 MHz to 18 GHz

True-RMS universal average power sensor covering 10 MHz to 18 GHz with an 80 dB dynamic range from -60 to +20 dBm.

Frequency Range
10 MHz - 18 GHz
Dynamic Range
-60 to +20 dBm
Detection
RMS
SWR (10-150 MHz)
<1.17

Overview

The Anritsu MA2482D is a member of the MA248xD series of universal power sensors for average power measurement. It covers 10 MHz to 18 GHz and offers a dynamic range of 80 dB, from -60 to +20 dBm. It can measure average power on CW, multi-tone and modulated RF waveforms such as 3G, 4G and OFDM signals.

The sensor architecture uses three pairs of diodes, each configured to operate in its square-law region over the sensor's dynamic range. As a result, the sensor measures true RMS power regardless of the type or bandwidth of the modulation on the input signal.

Option 1 (MA2480/01) adds a fast CW mode for high-speed measurement of CW signals, along with TDMA and pulse measurement capability. With this option, one of the diode pairs is calibrated for linearity over a wide dynamic range. Calibration factors are stored in the sensor's EEPROM.

Key Features

  • True-RMS detection for accurate average power measurements of any signal type
  • 80 dB dynamic range
  • Frequency coverage from 10 MHz to 18 GHz
  • Option 1 enables fast and accurate (1.8 % linearity) CW average power measurements
  • Option 1 adds TDMA and pulse measurement capability
  • Calibration factors stored in EEPROM
  • Precision RF connector with hexagon coupling nut for use with an industry-standard torque wrench

Specifications

RF Performance
Frequency Range 10 MHz - 18 GHz
Dynamic Range -60 to +20 dBm
Detection RMS
Listed as Video Bandwidth: RMS
SWR (10-150 MHz) <1.17
SWR (0.15-2 GHz) <1.12
SWR (2-12.4 GHz) <1.22
SWR (12.4-18 GHz) <1.25
Rise Time
MA2480/01
<0.004 ms
With option 1 only; 0.0 dBm, room temperature
Sensor Linearity (10 MHz to 6 GHz) 3% (-60 to +20 dBm)
Sensor Linearity (6 to 18 GHz) 3% (-60 to +20 dBm); 3.5% (0 to +20 dBm)
Manufacturer page prints the second range as 'o to +20 dBm'
Sensor Linearity (CW, Option 1)
MA2480/01
1.8% CW
With option 1
Physical
RF Connector N(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
Compliance
EMC Directive 2014/30/EU
Low Voltage Directive 2014/35/EU
RoHS Directive 2011/65/EU & 2015/863
Electromagnetic Compatibility EN 61326-1:2013
Emissions EN 55011:2009 +A1:2010 Group 1 Class A
Electrical Safety EN 61010-1:2010
Warranty
Warranty 1 Year

Applications

  • Average power measurement of CW signals
  • Multi-tone signal power measurement
  • Modulated RF signal measurement (3G, 4G, OFDM)
  • TDMA and pulse measurements (with Option 1)

Options & Accessories

Option Description
MA2480/01 Option 1: Fast CW mode, TDMA and pulse measurements — Adds fast CW mode to Universal Power Sensors for high speed measurements of CW signal plus TDMA and pulse measurements.

Documents & Resources