Anritsu MA247xD Series standard diode power sensors (CW), including the MA2473D
Anritsu / MA247xD / Diode Sensors
MA2473D

Anritsu MA2473D Standard Diode Power Sensor (CW), 10 MHz to 32 GHz

Standard CW diode power sensor covering 10 MHz to 32 GHz with a -70 dBm to +20 dBm dynamic range and K(m) connector.

Frequency Range
10 MHz - 32 GHz
Dynamic Range
-70 to +20
Rise Time
< 0.004
RF Connector
K(m)

Overview

The Anritsu MA2473D is a standard diode power sensor in the MA247xD series, covering 10 MHz to 32 GHz with a K(m) RF connector. It is designed for use with the ML2430, ML2480 and ML2490 series power meters.

The sensor is intended for high dynamic range, high accuracy CW and TDMA measurements. It provides a 90 dB dynamic range, from -70 dBm to +20 dBm, and stores its calibration factors in EEPROM.

With a rise time of 4 µs, the sensor is fast enough for power measurements on GSM and similar TDMA systems that use GMSK modulation. Its RF connector has a hexagon coupling nut for attachment with an industry standard torque wrench.

Key Features

  • Wide dynamic range: -70 dBm to +20 dBm
  • Accurate CW average power measurements
  • Frequency coverage from 10 MHz to 32 GHz
  • Calibration factors stored in EEPROM
  • 4 µs rise time for GSM and TDMA power measurements
  • Precision K(m) connector with hexagon coupling nut for torque wrench attachment

Specifications

General
Frequency Range 10 MHz - 32 GHz
Dynamic Range -70 to +20
Rise Time < 0.004
0.0 dBm, room temperature
RF Connector K(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
Compatible Power Meters ML2430, ML2480 and ML2490 series
SWR
SWR, 10-50 MHz <1.90
10-50 MHz
SWR, 50-150 MHz <1.17
50-150 MHz
SWR, 0.15-2 GHz <1.12
0.15-2 GHz
SWR, 2-12.4 GHz <1.22
2-12.4 GHz
SWR, 12.4-18 GHz <1.25
12.4-18 GHz
SWR, 18-32 GHz <1.35
18-32 GHz
Sensor Linearity
Sensor Linearity, < 18 GHz 1.8%
< 18 GHz
Sensor Linearity, < 40 GHz 2.5%
< 40 GHz
Compliance
EMC Directive 2014/30/EU
Low Voltage Directive 2014/35/EU
RoHS Directive 2011/65/EU & 2015/863
Electromagnetic Compatibility EN 61326-1:2013
Emissions EN 55011:2009 +A1:2010 Group 1 Class A
Electrical Safety EN 61010-1:2010
Warranty 1 Year Warranty

Applications

  • CW average power measurement
  • TDMA power measurement
  • GSM systems using GMSK modulation
  • Precision power measurement with ML2430, ML2480 and ML2490 series power meters