Anritsu MA247xD Series standard diode power sensors (CW), including the MA2472D
Anritsu / MA247xD / Diode Sensors
MA2472D

MA247xD Series MA2472D Standard Diode Power Sensor (CW), 10 MHz to 18 GHz

Standard CW diode power sensor covering 10 MHz to 18 GHz with a -70 dBm to +20 dBm dynamic range and N(m) connector.

Frequency Range
10 MHz - 18 GHz

Overview

The Anritsu MA2472D is a standard diode power sensor in the MA247xD Series. It is designed for use with the ML2430, ML2480 and ML2490 series power meters and covers the frequency range from 10 MHz to 18 GHz.

The sensor is intended for high dynamic range, high accuracy CW and TDMA measurements. It offers a 90 dB dynamic range, from -70 dBm to +20 dBm, with sensor linearity of 1.8% below 18 GHz.

With a rise time of less than 4 µs, the MA2472D is fast enough for power measurements on GSM and similar TDMA systems that use GMSK modulation. Calibration factors are stored in the sensor's EEPROM, and the RF input uses a precision N(m) connector with a hexagon coupling nut for attachment by an industry standard torque wrench.

Key Features

  • Wide dynamic range: -70 dBm to +20 dBm
  • Accurate CW average power measurements
  • Frequency coverage from 10 MHz to 18 GHz
  • Calibration factors stored in EEPROM
  • Rise time fast enough for GSM and similar TDMA systems using GMSK modulation
  • Compatible with ML2430, ML2480 and ML2490 series power meters

Specifications

Frequency
Frequency Range 10 MHz - 18 GHz
Amplitude
Dynamic Range -70 to +20 dBm
Sensor Linearity 1.8%, < 18 GHz
SWR
SWR, 10-150 MHz <1.17
MA2472D only
SWR, 0.15-2 GHz <1.12
SWR, 2-12.4 GHz <1.22
SWR, 12.4-18 GHz <1.25
Dynamics
Rise Time < 0.004 ms
0.0 dBm, room temperature
Video Bandwidth 100 kHz
Series listing summary
Physical
RF Connector N(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
Regulatory Compliance
EMC Directive 2014/30/EU
Low Voltage Directive 2014/35/EU
RoHS Directive 2011/65/EU & 2015/863
Electromagnetic Compatibility EN 61326-1:2013
Emissions EN 55011:2009 +A1:2010 Group 1 Class A
Electrical Safety EN 61010-1:2010
General
Warranty 1 Year

Applications

  • CW average power measurement
  • TDMA power measurement
  • GSM power measurement
  • Precision RF power measurement

Documents & Resources