Anritsu MA24400A series USB peak power sensors, including the MA24440A
Anritsu / MA24400A / USB Peak Power Sensors
MA24440A

Anritsu MA24440A Microwave Peak USB Power Sensor, 50 MHz to 40 GHz

USB microwave peak power sensor covering 50 MHz to 40 GHz with 70 MHz video bandwidth, 5 ns rise time and a K(m) connector.

Frequency Range
50 MHz to 40 GHz
Dynamic Range (Average)
-34 dBm to +20 dBm
Dynamic Range (Pulse)
-24 dBm to +20 dBm
Input Damage Level
+23 dBm, ±10.0 VDC (+30 dBm peak for 1 µs), minimum
Datasheet

Overview

The Anritsu MA24440A is a microwave peak power sensor in the MA24400A USB Peak Power Sensor family. It covers 50 MHz to 40 GHz and connects to a PC over USB. It is built to measure the peak power of wideband modulated signals and short RF pulses.

The sensor processes power readings in parallel and in real time. It does not pause measurement while it processes captured data, so pulses and glitches are captured with almost no gaps. It samples continuously at 100 MSa/s and has an effective sampling rate of 10 GSa/s. This gives fine time resolution for measuring rising and falling edges.

On the MA24440A, the fast video bandwidth is 70 MHz with a 5 ns rise time. The average-power measurement range is -34 dBm to +20 dBm. The RF input uses a ruggedized K male connector.

Software

The sensor ships with PC software for setting it up and viewing results. Display modes include peak and average power readouts, a trace view for pulse power analysis, and CCDF graphs. The software also supports crest factor and other statistical measurements. Multiple sensors can be synchronized for multi-channel measurements.

Key Features

  • 50 MHz to 40 GHz frequency range
  • 70 MHz video bandwidth with 5 ns rise time
  • 100,000 measurements per second
  • Real-time processing of power readings
  • 100 MSa/s continuous and 10 GSa/s effective sampling rates
  • 100 ps time resolution for rising/falling edge measurements
  • Full pulse profiling
  • Crest factor, CCDF and statistical measurements
  • Synchronized multi-channel measurements
  • Microwave Peak Power Analyzer measurement and analysis software

Specifications

Measurement Range
Frequency Range 50 MHz to 40 GHz
Dynamic Range (Average) -34 dBm to +20 dBm
Internal trigger
Dynamic Range (Pulse) -24 dBm to +20 dBm
Internal trigger
Input Damage Level +23 dBm, ±10.0 VDC (+30 dBm peak for 1 µs), minimum
Maximum Input 200 mW (+23 dBm) average, 1 W (+30 dBm) peak for 1 μs, ±10.0 VDC
VSWR
VSWR, 50 MHz to 4 GHz 1.25:1
Datasheet lists the next band as >2.0 GHz to 4 GHz; the band limits are printed as given
VSWR, >2.0 GHz to 4 GHz 1.40:1
VSWR, >4.0 GHz to 38 GHz 1.65:1
VSWR, >38 GHz to 40 GHz 2.00:1
Response
Sampling Rate 100 MSa/s
characteristic
Effective Sampling Rate 10 GSa/s
characteristic
Rise Time (Fast) 5 ns
Single channel
Rise Time (Standard) <10 µs
Single channel
Video Bandwidth (Fast) 70 MHz
Video Bandwidth (Standard) 350 kHz
Single Shot Bandwidth 35 MHz
Minimum Pulse Width (Fast) 10 ns
Dynamic range -10 dBm to +20 dBm
Minimum Pulse Width (Standard) 3 µs
Dynamic range -10 dBm to +20 dBm
Triggering
Source Continuous, Internal, External TTL, Crossover (from another sensor)
Arm Type Continuous, Single, Trigger Holdoff, Frame (gap) Holdoff
Slope Positive or Negative
Trigger Hold Off 100 ns to 1 s with 10 ns resolution
External Trigger
External Trigger Input SMB (female)
Impedance 10 kΩ
Type TTL
High Level Input Voltage 2.4 V min, 5.5 V max
Low Level Input Voltage -0.1 V min, 0.7 V max
Latency < 10 ns
Time delay between the defined edge of the applied trigger and the sensor switching into the triggered state
Trigger Pulse Width 10 ns min
Trigger Repetition Period 20 ns min
PC Requirements
Processor 1.3 GHz or higher recommended
RAM 512 MB (1 GB or more recommended)
Operating System Microsoft® Windows® 7 to 10 (32-bit and 64-bit)
Hard-Disk Free Space Min 1.0 GB free space to install and run
Display Resolution 800 x 600 (1280 x 1024 or higher recommended)
Interface USB 2.0 high speed
General
Interface to Host USB 2.0 high speed
RF Connector K male (ruggedized) to 40 GHz
Size 145 mm x 43 mm x 43 mm, excluding N connector
Series-wide value as printed in the datasheet
Weight 363 g (0.80 lb)
Warranty 1 year from the date of shipment
Warm-Up Time 60 minutes
Specification conditions
Calibration Cycle Anritsu recommended calibration interval is 12 months.
Connector Torque Tighten connector nut to 90 N·cm (8 lbf·in)
Do not rotate sensor body to connect or disconnect
Environmental
Test Standard Tests were performed per MIL-28800F (Class3).
Operating Temperature Range 0 ºC to 55 ºC
Storage Temperature Range -40 ºC to +70 ºC
Relative Humidity (non-condensing) 45 % at 50 ºC; 75 % at 40 ºC; 95 % at 30 ºC
Altitude 3048 m operational max
Shock 30 g half-sine, 11 ms duration
Vibration Sinusoidal: 5 Hz to 55 Hz, 3 g max; Random: 10 Hz to 500 Hz, 2.34 g rms; Power Spectral Density: 0.01 g2/Hz
Regulatory Compliance
EMC Directive 2014/30/EU, EN 61326:2013
European Union
Emissions CISPR 11/EN 55011:2019+A1:2010 Class A Equipment
Immunity IEC/EN 61000-4-2/3/4/6
Low Voltage Directive 2014/35/EU, EN 61010-1:2001
RoHS Directive 2015/863/EU

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • Radar systems
  • Pulsed communication
  • Aerospace and defense
  • Component testing
  • High-power microwave (HPM) applications
  • Wideband modulated signal measurement such as 802.11ac

Included

  • 11410-00976 Information Card
  • 806-391-R 1.8 m USB A (m) to USB B (m) cable

Options & Accessories

Option Description
MA24440A-097 Accredited Calibration — Accredited Calibration to ISO/IEC 17025 and ANSI/NCSL Z540-1 or ANSI/NCSL Z540.3 (includes test report, uncertainty data, and accreditation symbol)
41KB-3 Fixed Attenuator, 3 dB — Precision, DC to 26.5 GHz, 1W, 3 dB, K(m) to K(f)
41KB-6 Fixed Attenuator, 6 dB — Precision, DC to 26.5 GHz, 1W, 6 dB, K(m) to K(f)
41KB-10 Fixed Attenuator, 10 dB — Precision, DC to 26.5 GHz, 1W, 10 dB, K(m) to K(f)
41KB-20 Fixed Attenuator, 20 dB — Precision, DC to 26.5 GHz, 1W, 20 dB, K(m) to K(f)
43KC-3 Fixed Attenuator, 3 dB — Precision, DC to 26.5 GHz, 1W, 3 dB, K(m) to K(f)
43KC-6 Fixed Attenuator, 6 dB — Precision, DC to 26.5 GHz, 1W, 6 dB, K(m) to K(f)
43KC-10 Fixed Attenuator, 10 dB — Precision, DC to 26.5 GHz, 1W, 10 dB, K(m) to K(f)
43KC-20 Fixed Attenuator, 20 dB — Precision, DC to 26.5 GHz, 1W, 20 dB, K(m) to K(f)
41KC-3 Fixed Attenuator, 3 dB — Precision, DC to 40 GHz, 1W, 3 dB, K(m) to K(f)
41KC-6 Fixed Attenuator, 6 dB — Precision, DC to 40 GHz, 1W, 6 dB, K(m) to K(f)
41KC-10 Fixed Attenuator, 10 dB — Precision, DC to 40 GHz, 1W, 10 dB, K(m) to K(f)
41KC-20 Fixed Attenuator, 20 dB — Precision, DC to 40 GHz, 1W, 20 dB, K(m) to K(f)
34NFK50 Precision Adapter — Precision Adapter, DC to 18 GHz, N(f) to K(m), 50Ω
34NK50 Precision Adapter — Precision Adapter, DC to 18 GHz, N(m) to K(m), 50Ω
34NKF50 Precision Adapter — Precision Adapter, DC to 18 GHz, N(m) to K(f), 50Ω

Documents & Resources