Anritsu MA2400xA series thermal power sensors with N, K and V coaxial connectors
Anritsu / Series · Power Sensors

MA2400xA Series Thermal Power Sensors, 10 MHz to 50 GHz

Thermal power sensors with true-RMS detection and 50 dB dynamic range, 10 MHz to 50 GHz, for Anritsu ML2430A, ML2480 and ML2490 series power meters.

Frequency Range
10 MHz to 18 GHz
Frequency Range
10 MHz to 40 GHz
Frequency Range
10 MHz to 50 GHz
View Models Datasheet

Overview

The Anritsu MA2400xA series thermal power sensors are designed for use with the ML2430, ML2480 and ML2490 series power meters. The series comprises three models that share the same –30 dBm to +20 dBm measurement range and differ in upper frequency limit and RF connector: the MA24002A covers 10 MHz to 18 GHz with an N(m) connector, the MA24004A covers 10 MHz to 40 GHz with a K(m) connector, and the MA24005A covers 10 MHz to 50 GHz with a V(m) connector.

Measurement principle

Thermal sensors use Seebeck elements. The combined effect of a thermal gradient and charge migration between dissimilar metals produces a true reading of the average power of any incident waveform, independent of modulation type or access scheme. This true-RMS detection gives accurate average power measurements over 50 dB of dynamic range.

Calibration and connectors

Each sensor stores its calibration factor and linearity correction data in a built-in EEPROM, so the power meter applies the corrections automatically. The sensors incorporate precision RF connectors with a hexagon coupling nut for attachment with an industry-standard torque wrench, which helps maintain consistent, repeatable connections and protects the connector from damage.

Models in this series

Model Description
MA24002A
Anritsu MA24002A Thermal Power Sensor, 10 MHz to 18 GHz
Thermal power sensor covering 10 MHz to 18 GHz with -30 to +20 dBm dynamic range and an N(m) connector for Anritsu power meters.
Details →
MA24004A
Anritsu MA24004A Thermal Power Sensor, 10 MHz to 40 GHz
Thermal power sensor covering 10 MHz to 40 GHz with a –30 dBm to +20 dBm dynamic range and true-RMS average power measurement.
Details →
MA24005A
Anritsu MA24005A Thermal Power Sensor, 10 MHz to 50 GHz
Thermal power sensor covering 10 MHz to 50 GHz with true average power detection over a –30 dBm to +20 dBm dynamic range and a V(m) connector.
Details →

Key Features

  • True-RMS detection enables accurate average power measurements of any signal type
  • 50 dB dynamic range
  • Wide frequency coverage from 10 MHz to 50 GHz (sensor dependent)
  • Calibration factors stored in EEPROM
  • Low SWR across the frequency range
  • Precision RF connectors with hexagon coupling nut for torque-wrench attachment

Specifications

Frequency
Frequency Range
MA24002A
10 MHz to 18 GHz
Frequency Range
MA24004A
10 MHz to 40 GHz
Frequency Range
MA24005A
10 MHz to 50 GHz
Power
CW Dynamic Range –30 dBm to +20 dBm
Sensor Linearity, < 18 GHz 1.8%
± value. Linearity applicable from –30 to +15 dBm; add 1% for power levels > +15 dBm
Sensor Linearity, < 40 GHz 2.0%
± value. Linearity applicable from –30 to +15 dBm; add 1% for power levels > +15 dBm
Sensor Linearity, < 50 GHz 2.5%
± value. Linearity applicable from –30 to +15 dBm; add 1% for power levels > +15 dBm
Temperature Accuracy < 1% < 30 GHz < +10 dBm, < 1.5% ≥30 GHz ≥ +10 dBm
Rise Time < 15 ms
0.0 dBm, room temperature with standard 1.5 m sensor cable
Power Measurement Average (RMS)
RF Input
SWR, 10 MHz to 50 MHz < 1.90
Nominal input impedance is 50 ohms
SWR, 50 MHz to 150 MHz < 1.17
SWR, 150 MHz to 2 GHz < 1.10
SWR, 2 GHz to 12.4 GHz < 1.15
SWR, 12.4 GHz to 18 GHz < 1.20
SWR, 18 GHz to 32 GHz
MA24004A, MA24005A
< 1.25
SWR, 32 GHz to 40 GHz
MA24004A, MA24005A
< 1.30
SWR, 40 GHz to 50 GHz
MA24005A
< 1.40
Input Impedance 50 ohms
Nominal
RF Connector
MA24002A
N(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
RF Connector
MA24004A
K(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
RF Connector
MA24005A
V(m)
Precision RF connector with hexagon coupling nut for attachment by industry standard torque wrench
General
Compatible Power Meters ML24xxA/B
Page lists ML2430, ML2480 and ML2490 series power meters
Measurement Application (Examples) Any Modulation; Any Access Scheme
Calibration Data Storage Calibration factors stored in EEPROM
Recommended Calibration Cycle 12 months
Warranty 1 year
Standard warranty for power sensors
Regulatory Compliance
EMC Directive 2014/30/EU, EN 61326-1:2013
Emissions EN 55011:2009 +A1:2010 Group 1 Class A
Low Voltage Directive 2014/35/EU
Electrical Safety EN 61010-1:2010
RoHS Directive 2011/65/EU & 2015/863
Australia and New Zealand RCM AS/NZS 4417:2012
South Korea KCC-REM-A21+-0004
Canada ICES-1(A)/NMB-1(A)

Applications

  • Average power measurement of signals with any modulation type
  • Power measurement for any access scheme
  • CW and modulated signal power measurement with ML2430A, ML2480 and ML2490 series power meters

Options & Accessories

Option Description
MA24000-097 Option 97, Accredited Calibration — For MA2400XA Thermal Sensors
MA24000-098 Option 98, Standard Calibration — For MA2400XA Thermal Sensors
MA24000-099 Option 99, Premium Calibration — For MA2400XA Thermal Sensors
2000-1536-R 0.3 m Sensor Cable
2000-1537-R 1.5 m Sensor Cable — Standard sensor cable; specifications apply with this cable
2000-1538-R 3 m Sensor Cable
2000-1539-R 5 m Sensor Cable
2000-1540-R 10 m Sensor Cable
2000-1541-R 30 m Sensor Cable
2000-1542-R 50 m Sensor Cable
2000-1543-R 100 m Sensor Cable
D41310 Soft Carry Case with Shoulder Strap

Documents & Resources